| APPENDIX | M300EU Addendum to M300E/EM Manual | ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| SETUP VARIABLE |
|
| NUMERIC |
|
|
| DEFAULT |
|
|
| VALUE RANGE |
| DESCRIPTION | |
|
|
|
| UNITS |
|
|
| VALUE |
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
| O2_FILT_PCT | % |
| 2 |
|
|
|
| Relative change in O2 concentration to | ||||||
|
|
|
|
|
|
| shorten filter. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| O2_FILT_DELAY |
| Seconds | 20 |
|
|
|
| Delay before leaving O2 adaptive filter mode. | ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| O2_DIL_FACTOR |
| — | 1 |
|
|
|
| Dilution factor for O2. Used only if is dilution | ||||||
|
|
|
|
|
|
| enabled with FACTORY_OPT variable. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| CO2_COMP_ENABLE |
| — |
|
| OFF |
|
| ON, OFF | ON enables CO2 compensation; OFF disables | |||||
|
|
|
|
|
|
| it. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| CO2_COMP_CONC | % |
| 0 |
|
|
|
| CO2 concentration to compensate for. | ||||||
|
| O2_CELL_SET |
| ºC |
| 50 |
|
|
|
| O2 sensor cell temperature set point and | |||||
|
|
|
|
| Warnings: |
|
| |||||||||
|
|
|
|
|
|
| warning limits. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
| STD_O2_CELL_TEMP |
| ºK | 323 |
|
|
|
| Standard O2 cell temperature for temperature | ||||||
|
|
|
|
|
|
| compensation. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| ZERO_ENABLE |
| — |
|
| ON, OFF |
|
| OFF, ON | ON enables | |||||
|
|
|
|
|
|
| scrubber; OFF disables it. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| ZERO_FREQ |
| Minutes | 5 |
|
|
|
| |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| CO2_TARG_SPAN_CONC | % |
| 12 |
|
|
|
| Target CO2 concentration during span | ||||||
|
|
|
|
|
|
| calibration. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| CO2_SLOPE |
| — | 1 |
|
|
|
| CO2 slope. | ||||||
|
| CO2_OFFSET | % |
| 0 |
|
|
|
| CO2 offset. | ||||||
|
| O2_TARG_SPAN_CONC | % |
| 20.95 |
|
|
|
| Target O2 concentration during span | ||||||
|
|
|
|
|
|
| calibration. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| O2_SLOPE |
| — | 1 |
|
|
|
| O2 slope. | ||||||
|
|
|
|
|
|
|
|
|
|
|
| |||||
|
| O2_OFFSET | % |
| 0 |
|
|
|
| O2 offset. | ||||||
|
| CO2_RANGE | % |
| 15 |
|
|
|
| CO2 concentration range. | ||||||
|
| O2_RANGE | % |
| 100 |
|
|
|
| O2 concentration range. | ||||||
|
| O2_CELL_CYCLE |
| Seconds | 10 |
|
|
|
| O2 cell temperature control cycle period. | ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| O2_CELL_PROP |
| — | 1 |
|
|
|
| O2 cell PID temperature control proportional | ||||||
|
|
|
|
|
|
| coefficient. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| O2_CELL_INTEG |
| — | 0.1 |
|
|
|
| O2 cell PID temperature control integral | ||||||
|
|
|
|
|
|
| coefficient. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| O2_CELL_DERIV |
| — |
|
| 0 (disabled) |
|
| O2 cell PID temperature control derivative | ||||||
|
|
|
|
|
|
| coefficient. | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 05924 Rev K.6 |