| APPENDIX | M300EU Addendum to M300E/EM Manual |
| ||||||
|
| Table | Additional or Changed Test Function Relevant to the M300EU , Revision K.6 |
| |||||
|
|
|
|
|
|
|
|
|
|
|
| TEST FUNCTION |
|
| MESSAGE TEXT |
|
| DESCRIPTION |
|
|
| NAME |
|
|
|
|
| ||
|
|
|
|
|
|
|
|
| |
|
| STABILITY |
|
| STABIL=0.0 PPM | These gases not applicable to the M300EU |
| ||
|
|
|
|
|
|
|
| ||
|
| AUTOZERO |
|
| AZERO RATIO=1.234 | Measure/reference ratio during |
| ||
|
|
|
|
|
|
|
| ||
|
| BOXTEMP2 |
|
| BOX TEMP2=29.6 C | Internal box temperature #2. |
| ||
|
|
|
|
|
|
|
| ||
|
| OVENTEMP |
|
| OVEN TEMP=30.1 C | Oven temperature |
| ||
|
|
|
|
|
|
|
|
|
|
Table
Because the CO2 and O2 sensor options are not available on the M300EU, the following test functions listed in Appendix
TEST FUNCTION |
| MESSAGE TEXT |
|
| DESCRIPTION |
|
|
|
|
| |||
NAME |
|
|
|
| ||
|
|
|
|
|
| |
CO2RANGE |
| CO2 RANGE=20 % 1 | CO2 | range. | ||
O2RANGE |
| O2 RANGE=100 % 2 | O2 range. | |||
|
| CO STB=0.0 PPM 1, 2 |
|
|
| |
[STABILITY] |
| CO2 STB=0.0 % 1 | These gases not applicable to the M300EU | |||
|
| O2 STB=0.0 % 2 |
|
|
| |
O2CELLTEMP |
| O2 CELL TEMP=50.2 C | O2 sensor cell temperature. | |||
|
|
|
|
| ||
CO2SLOPE |
| CO2 SLOPE=1.000 | CO2 | slope, computed during zero/span calibration. | ||
|
|
|
|
| ||
CO2OFFSET |
| CO2 OFFSET=0.000 | CO2 | offset, computed during zero/span calibration. | ||
|
|
|
| |||
O2SLOPE |
| O2 SLOPE=0.980 | O2 slope, computed during zero/span calibration. | |||
|
|
|
| |||
O2OFFSET |
| O2 OFFSET=1.79 % | O2 offset, computed during zero/span calibration. | |||
|
|
|
|
|
|
|
| 05924 Rev K.6 |