M300EU | M300EU OPERATING INSTRUCTIONS |
(Addendum to M300EU Manual - P/N 04145) |
|
6.7.2. CALIBRATION OF THE ANALOG OUTPUTS
This section replaces Section 6.13.6 of the M300E/EM Operators Manual (P/N 04288).
ANALOG OUPUT calibration needs to be carried out on first startup of the analyzer (performed in the factory as part of the configuration process) or whenever
During automatic calibration, the analyzer tells the output circuitry to generate a zero mV signal and
Automatic calibration can be performed via the AOUTS CALIBRATION command, or by using the CAL button located inside each of the output submenus (e.g. CONC_OUT_1, CONC_OUT_2 or TEST_OUTPUT). By default, the analyzer is configured so that calibration of analog outputs can be initiated as a group with the AOUT CALIBRATION command or individually.
ENABLING OR DISABLING THE AUTO-CAL FOR AN INDIVIDUAL ANALOG OUTPUT
The information found in Section 6.13.5.9 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):
•There are minor differences in the displays ( e.g. CONC_OUT_2 instead of DATA_OUT_2)
AUTOMATIC ANALOG OUTPUT CALIBRATION
The information found in Section 6.13.6.2 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU.
MANUAL CALIBRATION OF ANALOG OUTPUT CONFIGURED FOR VOLTAGE RANGES
The information found in Section 6.13.6.3 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):
•There are minor differences in the displays ( e.g. CONC_OUT_2 instead of DATA_OUT_2) User Notes:
MANUAL CALIBRATION OF ANALOG OUTPUTS CONFIGURED FOR CURRENT LOOP RANGES
The information found in Section 6.13.6.4 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s):
•There are minor differences in the displays ( e.g. CONC_OUT_2 instead of DATA_OUT_2)
•The current output option can only be installed on outputs A1 and A2 of the M300EU
CALIBRATING THE INTERNAL A TO D CONVERTER (AIN CALIBRATION)
The information found in Section 6.13.6.5 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU.
05514 Rev A1 | 37 |