Teledyne Model 300EU Analog Output Configuration, Analog I/O Configuration, M300EU, Sub Menu

Page 42
6.7. ANALOG OUTPUT CONFIGURATION

M300EU OPERATING INSTRUCTIONS

M300EU

 

Addendum to M300E/EM Manual P/N 04288

6.7. ANALOG OUTPUT CONFIGURATION

6.7.1. ANALOG I/O CONFIGURATION

Because the manner in which the analog outputs are implemented differs between the M300E/EM and the M300EU, there are significant differences in the manner in which the ANALOG I/O are manipulated.

THE ANALOG I/O CONFIGURATION SUBMENU

This section replaces Section 6.13.5.1 of the M300E/EM Operators Manual (P/N 04288).

 

Table 6-5:

DIAG - Analog I/O Functions

 

 

 

 

SUB MENU

 

 

FUNCTION

 

Initiates a calibration of the A1, A2 and A4 analog output channels that

AOUT CALIBRATED

determines the slope and offset inherent in the circuitry of each output.

These values are stored in the and applied to the output signals by the CPU

 

 

automatically

 

 

 

 

Sets the basic electronic configuration of the A1 output. There are four

 

options:

 

 

RANGE: Selects the signal type (voltage or current loop) and level of the

 

 

output

 

CONC_OUT_11

A1 OFS: Allows them input of a DC offset to let the user manually adjust

 

 

the output level

 

AUTO CAL: Enables / Disables the AOUT CALIBRATION Feature

 

CALIBRATED: Performs the same calibration as AOUT CALIBRATED,

 

 

but on this one channel only.

 

Sets the basic electronic configuration of the A2 output. There are three

 

options:

 

 

RANGE: Selects the signal type (voltage or current loop) and level of the

 

 

output

 

CONC_OUT_21

A2 OFS: Allows them input of a DC offset to let the user manually adjust

 

 

the output level

 

AUTO CAL: Enables / Disables the AOUT CALIBRATION Feature

 

CALIBRATED: Performs the same calibration as AOUT CALIBRATED,

 

 

but on this one channel only.

 

Sets the basic electronic configuration of the A4 output. There are three

 

options:

 

 

RANGE: Selects the signal type (voltage or current loop) and level of the

 

 

output

 

TEST OUTPUT1

A4 OFS: Allows them input of a DC offset to let the user manually adjust

 

 

the output level

 

AUTO CAL: Enables / Disables the AOUT CALIBRATION Feature

 

CALIBRATED: Performs the same calibration as AOUT CALIBRATED,

 

 

but on this one channel only.

AIN CALIBRATED

Initiates a calibration of the A-to-D Converter circuit located on the Mother

Board.

 

 

 

 

 

1Changes to RANGE or REC_OFS require recalibration of this output.

34

05831 Rev. 01

Image 42
Contents TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION TAPI ADDENDUM TO M300E/EM MANUAL P/NMODEL 300EU CO ANALYZER WITH AUTO-REFERENCE 9480 CARROLL PARK DRIVE SAN DIEGO, CALIFORNIA USAPage M300EU SAFETY MESSAGESUSER NOTES GENERAL SAFETY HAZARDAddendum to M300E/EM Manual P/N TABLE OF CONTENTS LIST OF FIGURES LIST OF TABLESLIST OF APPENDICES APPENDIX A-4 Model 300EU Signal I/O Definitions, Revision K.605904 Rev A 1.2. M300EU OVERVIEW 1. INTRODUCTION1.1. REFERENCE NUMBERING CONVENTION any time the instrument is in SETUP mode05904 Rev A M300EU Basic Unit Specifications 2. SPECIFICATIONS, APPROVALS AND WARRANTY2.1. SPECIFICATIONS 05514 Rev A12.3. CE MARK COMPLIANCE 2.4. WARRANTY2.2. EPA EQUIVALENCY DESIGNATION 05831 Rev3.2. M300EU ANALYZER LAYOUT 3. GETTING STARTED3.1. UNPACKING THE M300EU OPTICAL BENCHM300EU 3.3.1. POWER CONNECTION3.3. ELECTRICAL CONNECTIONS Figure 3-2 M300EU Internal Gas Flow3.3.2. ANALOG OUTPUT Connections Analog Output Data Type Default SettingsFigure 3-3 M300EU Analog Output Connector Table 3-3 Status Output Pin Assignments 3.3.3. CONNECTING THE STATUS OUTPUTSAnalog Output Pin Outs Output #3.4.2. WARNING MESSAGES 3.4. INITIAL OPERATION OF THE M300EU3.4.1. STARTUP AND WARM UP OF THE M300EU MESSAGE3.4.3. FUNCTIONAL CHECK The Test functions available from the front panel of the M300EU are3.5.1. PRE-CALIBRATION STABILIZATION 3.5.2. BASIC CO CALIBRATION SETUP3.5. INITIAL CALIBRATION 3.5.2.1. Calibration GasesSTEP 2 - DILUTION RATIO 3.5.3. BASIC CO CALIBRATION PROCEDURESTEP 1 - SET LIMITS STEP 3 - SET CO SPAN GAS CONCENTRATION4. FREQENTLY ASKED QUESTIONS & GLOSSARY 05831 Rev DESCRIPTION 5. OPTIONAL HARDWARE AND SOFTWAREOPTION 300ETOLL-FREE FAX TEL E-MAIL api-sales@teledyne.com 5.1. CALIBRATION VALVE OPTIONS FOR THE M300EU5.1.1. ZERO SPAN VALVES OPT 50A M300EU Internal Pneumatic Flow- Zero/Span Valves OPT 50A5.1.2. ZERO SPAN VALVES WITH IZS OPT 51B SAMPLE CHAMBERScrubber 5.1.3. Zero Span Valves with IZS and Shutoff VALVE OPT 51B 6. M300EU OPERATING INSTRUCTIONS sample gas stream is being routed through the CO scrubber of the auto 6.2. OPERATING MODESM300EU Operating Modes reference gas pathMode Field 6.2.1. AUTO-REFERENCE MODE AREFSETUP AUTO-REF RANGE = 50.000 PPMTable 6-2 Test Functions Defined 6.2.2. TEST FUNCTIONSUnits Parameter6.3.2. ANALOG OUTPUT RANGES FOR CO CONCENTRATION 6.3. SETUP Æ RNGE ANALOG OUTPUT REPORTING RANGE CONFIGURATION6.3.1. PHYSICAL RANGE VERSUS ANALOG OUTPUT REPORTING RANGES ANALOG OUTDUAL DEFAULT SETTINGSSNGL AUTOThis is the default reporting range mode for the analyzer To set the ranges press following keystroke sequence M300EU OPERATING INSTRUCTIONS Pressure STP 6.3.6. SETUP Æ RNGE Æ UNIT SETTING THE REPORTING RANGE UNIT TYPEweight/volume units CO ppb x 1.25 = CO ug/m3 CO ppm x 1.25= CO mg/m3 Consult your local regulations for the STP used by your agency6.4.1. ADJUSTING THE A-REF MODE CYCLE TIME 6.4.2. FORCING AN AUTO-REFERENCE MEASUREMENT To manually start an auto-reference measurement cycle, pressM300EU VARIABLE 6.5. SETUP Æ MORE Æ VARS INTERNAL VARIABLES VARSTable 6-3 Test Functions Defined DESCRIPTIONDIAG AOUT 6.6. SETUP Æ MORE Æ DIAG USING THE DIAGNOSTICS FUNCTIONSDIAG I/O DIAG AIOTHE ANALOG I/O CONFIGURATION SUBMENU 6.7. ANALOG OUTPUT CONFIGURATION6.7.1. ANALOG I/O CONFIGURATION DIAG - Analog I/O Functions4. Choose a Test Channel parameter to be output on the channel 6.13.5.7 SETTING THE REPORTING RANGE SCALE FOR AN ANALOG OUTPUT ANALOG OUTPUT SIGNAL TYPE AND RANGE SELECTIONTURNING THE ANALOG OUTPUT OVER-RANGE FEATURE ON/OFF 6.13.5.8 SETTING DATA UPDATE RATE FOR AN ANALOG OUTPUTENABLING OR DISABLING THE AUTO-CAL FOR AN INDIVIDUAL ANALOG OUTPUT 6.7.2. CALIBRATION OF THE ANALOG OUTPUTSMANUAL CALIBRATION OF ANALOG OUTPUT CONFIGURED FOR VOLTAGE RANGES AUTOMATIC ANALOG OUTPUT CALIBRATIONTEST CHANNEL 6.7.3. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4Test Channels Functions Available on the M300EU’s Analog Output ZERO6.7.4. HESSEN PROTOCOL Table 6-7 M300EU Hessen GAS ID ListDefault Hessen Status Bit Assignments REQUIRED EQUIPMENT, SUPPLIES AND EXPENDABLES 7. CALIBRATION PROCEDURESmanual for troubleshooting tips MANUAL CALIBRATION STEP3 - ZERO/SPAN CALIBRATIONMANUAL CALIBRATION WITH ZERO/SPAN VALVES 7.9.1 SECTION O2 SENSOR CALIBRATION PROCEDURE7.9.2 CO2 SENSOR CALIBRATION PROCEDURE MANUAL CALIBRATION CHECKS WITH ZERO/SPAN VALVES8. EPA PROTOCOL CALIBRATION 05831 Rev 9. INSTRUMENT MAINTENANCE 9.1. PERFOMING AN AUTO-REFERENCE MEASUREMENT AFTER MAINTENANCE05831 Rev CO MEAS Table 9-1 M300EU Test Function RecordSTABILITY MR RATIO05831 Rev 10.1.1. AUTO-REFERENCE RATIO 10. THEORY OF OPERATION10.1. MEASUREMENT METHOD Figure 10-1 M300EU Gas flow during Auto-Reference MeasurementsFigure 10-2 Auto-Reference Measurement Cycle 10.2. PNEUMATIC OPERATION10.2.1. THE A-REF CARBON MONOXIDE SCRUBBER This includes calibration gases such as zero air and span gasAuto-Reference Valve Operating States 10.2.2. THE A-REF VALVE ASSEMBLY10.2.3. THE NAFION→ DRYER Figure 10-3 Semi-Permeable Membrane Drying ProcessThe dryer consists of a bundle of 0.030 O.D. Nafion→ tubes housed within a single large tube shell. Sample gas flows within the Nafion tubes while water vapor absorbs into the tubular membrane walls and is removed. Moisture permeating the tubing is carried away by a dry purge gas within the shell that flows countercurrent to the sample gas Figure 10-4 M300EU Electronic Overview Block Diagram 10.3. ELECTRONIC OPERATION10.3.1. OVERVIEW Aout10.3.2. THE RELAY PCA 10.3.2.1. Temperature Control of the Convection OvenFigure 10-5 M300EU Heating Control Block Diagram 230 VAC Configuration 10.3.2.2. Oven Heaters AC Power Configuration115 VAC Configuration Figure 10-6 M300EU Oven Heater Configuration Jumpers10.3.3.1. A to D Conversion 10.3.3. MOTHERBOARD10.3.2.3. Status LED’s 10.3.3.2. Sensor Inputs10.3.4. POWER DISTRIBUTION RELAY PCAFigure 10-8 M300EU Distribution Block Diagram Page MAINTENANCE PERSONNEL ONLY 11. TROUBLESHOOTING & REPAIR11.1. GENERAL NOTES PROCEDURES11.2. FAULT DIAGNOSIS WITH WARNING MESSAGES Warning Messages - Indicated FailuresTable continued on next page Warning Messages - Indicated Failures cont Sample flow rate istable continued 11.2.1. FAULT DIAGNOSIS WITH TEST FUNCTIONSTable 11-2 Test Functions - Indicated Failures INDICATED FAILURESTest Functions - Indicated Failures Values outside range indicateTable 11-3 Relay Board Status LED Failure Indications 11.2.2. RELAY BOARD STATUS LED’SFaulty Connectors/Wiring Voltage displayed should change. If not11.3. GAS FLOW PROBLEMS 11.4. OTHER PERFORMANCE PROBLEMS11.4.1. Unexplained DRIFT 11.5.2. MOTHERBOARD 11.5. SUBSYSTEM CHECHOUT11.5.1. RELAY BOARD 11.5.2.1. A/D Functions11.6. TECHNICAL ASSISTANCE 05514 Rev A1 APPENDIX A-4 M300EU Signal I/O Definitions APPENDIX A - Version Specific Software DocumentationAPPENDIX A-2 M300EU Setup Variables Available Via Serial I/O M300EU Addendum to M300E/EM ManualCLR1 TEST1MSG1 Figure A-2 Primary Setup Menu Except iDAS Figure A-3 Primary Setup Menu Æ iDAS Submenu Figure A-4 Secondary Setup Menu Æ COMM and VARS Submenus Figure A-5 Secondary Setup Menu Æ Hessen Protocol Submenu Figure A-6 DIAG Menu ENTER PASSWORDAuto Cal feature is OFF or the range is set for Press ENTRAPPENDIX A-2 Setup Variables For Serial I/O, Revision K.6 APPENDIX A-2 Setup Variables For Serial I/O, Revision K.6Table continues CO2FILTPCT A-10 O2FILTPCTSTDO2CELLTEMP MESSAGE TEXT APPENDIX A-3 Warnings and Test Functions, Revision K.6NAME A-11These gases not applicable to the M300EU Table A-5A-12 These gases not applicable to the M300EUDescription Signal NameBit or Channel NumberAPPENDIX A-5 M300EU iDAS Parameters Revision K.6 OVTMPW Internal box temperature #2/oven warningA-14 O2SLPE A-15CO2OFS O2OFSTAPPENDIX A-6 Terminal Command Designators, Revision K.6 A-16European standard EN APPENDIX B - M300EU SPARE PARTS & EXPENDABLESAPPENDIX B - M300EU SPARE PARTS & EXPENDABLES 05925 Rev AM300EU Addendum to M300E/EM Manual