M300EU | TROUBLESHOOTING & REPAIR |
(Addendum to M300EU Manual - P/N 04145) |
|
Table | Test Functions - Indicated Failures |
TEST
FUNCTIONS
(As Displayed)
PHT DRIVE
SLOPE
OFFSET
INDICATED FAILURE(S)
If this drive voltage is out of range it may indicate one of several problems:
-A poor mechanical connection between the various components in inside the detector housing
-An electronic failure of the IR
-A temperature problem inside the analyzer chassis. In this case other temperature warnings would also be active such as OVEN TEMP WARNING, BENCH TEMP WARNING or BOX TEMP WARNING.
Values outside range indicate
Contamination of the zero air or span gas supply Instrument is Miscalibrated
Blocked gas flow
Contaminated or leaking GFC wheel (either chamber) Faulty IR
Faulty sample faulty IR
Bad/incorrect span gas concentration due. Values outside range indicate Contamination of the zero air supply Contaminated or leaking GFC wheel (either chamber) Faulty IR
05514 Rev A1 | 63 |