M300EU | M300EU OPERATING INSTRUCTIONS |
(Addendum to M300EU Manual - P/N 04145) |
|
To configure the analyzer’s three analog outputs, set the electronic signal type of each channel and calibrate the outputs. This consists of:
1.Selecting an output type (voltage or current, if an optional current output driver has been installed) and the signal level that matches the input requirements of the recording device attached to the channel.
2.Determine if the
3.If a Voltage scale is in use, a bipolar recorder offset may be added to the signal if required.
4.Choose a Test Channel parameter to be output on the channel.
5.Calibrating the output channel. This can be done automatically or manually for each channel.
To access the analog I/O configuration sub menu, press:
SAMPLE | RANGE=50.000 PPM | CO= XXXX | ||||
<TST | TST> | CAL | SETUP | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SETUP X.X | PRIMARY SETUP MENU |
|
| |||
CFG | DAS | RNGE PASS CLK MORE | EXIT | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Toggle these
keys to enter the
correct
PASSWORD
| SETUP X.X | SECONDARY SETUP MENU | |||||||||||
| COMM | VARS DIAG AREF |
| EXIT | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
| |||
| SETUP X.X | ENTER PASSWORD:818 |
|
|
| ||||||||
|
|
| 8 | 1 | 8 |
|
|
|
| ENTR EXIT | |||
|
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| DIAG |
|
|
| SIGNAL I/O |
|
|
| |||||
|
|
|
| NEXT |
|
|
|
| ENTR |
| EXIT |
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Continue pressing NEXT until ...
AIO Configuration Submenu
DIAG |
| ANALOG I/O CONFIGURATION |
| |||
PREV | NEXT |
| ENTR | EXIT | ||
|
|
|
|
|
| |
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DIAG AIO | A OUTS CALIBRATED: NO |
| ||||
<SET | SET> | CAL | EXIT | |||
|
|
|
|
| ||
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DIAG AIO | CONC_OUT_1: 5V, OVR, NOCAL | |||||
<SET | SET> | EDIT | EXIT | |||
|
|
|
|
| ||
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DIAG AIO | CONC_OUT_2: 5V, OVR, NOCAL | |||||
<SET | SET> | EDIT | EXIT | |||
|
|
|
|
| ||
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DIAG AIO | TEST_OUTPUT: 5V,OVR, NOCAL | |||||
<SET | SET> | EDIT | EXIT | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
DIAG AIO | AIN CALIBRATED: NO |
| ||||
<SET | SET> | CAL | EXIT | |||
|
|
|
|
|
|
|
Adjusts the signal output
for Analog Output A1
Adjusts the signal output
for Analog Output A2
Selects the parameter top be output on the TEST channel and adjusts its signal output
05514 Rev A1 | 35 |