Teledyne Model 300EU M300EU, Choose a Test Channel parameter to be output on the channel, Rev A1

Page 43
4.Choose a Test Channel parameter to be output on the channel.

M300EU

M300EU OPERATING INSTRUCTIONS

(Addendum to M300EU Manual - P/N 04145)

 

To configure the analyzer’s three analog outputs, set the electronic signal type of each channel and calibrate the outputs. This consists of:

1.Selecting an output type (voltage or current, if an optional current output driver has been installed) and the signal level that matches the input requirements of the recording device attached to the channel.

2.Determine if the over-range feature is needed and turn it on or off accordingly.

3.If a Voltage scale is in use, a bipolar recorder offset may be added to the signal if required.

4.Choose a Test Channel parameter to be output on the channel.

5.Calibrating the output channel. This can be done automatically or manually for each channel.

To access the analog I/O configuration sub menu, press:

SAMPLE

RANGE=50.000 PPM

CO= XXXX

<TST

TST>

CAL

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

PRIMARY SETUP MENU

 

 

CFG

DAS

RNGE PASS CLK MORE

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Toggle these

keys to enter the

correct

PASSWORD

 

SETUP X.X

SECONDARY SETUP MENU

 

COMM

VARS DIAG AREF

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

ENTER PASSWORD:818

 

 

 

 

 

 

8

1

8

 

 

 

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG

 

 

 

SIGNAL I/O

 

 

 

 

 

 

 

NEXT

 

 

 

 

ENTR

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Continue pressing NEXT until ...

AIO Configuration Submenu

DIAG

 

ANALOG I/O CONFIGURATION

 

PREV

NEXT

 

ENTR

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

A OUTS CALIBRATED: NO

 

<SET

SET>

CAL

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

CONC_OUT_1: 5V, OVR, NOCAL

<SET

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

CONC_OUT_2: 5V, OVR, NOCAL

<SET

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

TEST_OUTPUT: 5V,OVR, NOCAL

<SET

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

AIN CALIBRATED: NO

 

<SET

SET>

CAL

EXIT

 

 

 

 

 

 

 

Adjusts the signal output

for Analog Output A1

Adjusts the signal output

for Analog Output A2

Selects the parameter top be output on the TEST channel and adjusts its signal output

05514 Rev A1

35

Image 43
Contents 9480 CARROLL PARK DRIVE SAN DIEGO, CALIFORNIA USA ADDENDUM TO M300E/EM MANUAL P/NMODEL 300EU CO ANALYZER WITH AUTO-REFERENCE TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION TAPIPage GENERAL SAFETY HAZARD SAFETY MESSAGESUSER NOTES M300EUAddendum to M300E/EM Manual P/N TABLE OF CONTENTS LIST OF TABLES LIST OF FIGURESAPPENDIX A-4 Model 300EU Signal I/O Definitions, Revision K.6 LIST OF APPENDICESAddendum to M300E/EM Manual P/N any time the instrument is in SETUP mode 1. INTRODUCTION1.1. REFERENCE NUMBERING CONVENTION 1.2. M300EU OVERVIEWAddendum to M300E/EM Manual P/N 05514 Rev A1 2. SPECIFICATIONS, APPROVALS AND WARRANTY2.1. SPECIFICATIONS M300EU Basic Unit Specifications05831 Rev 2.4. WARRANTY2.2. EPA EQUIVALENCY DESIGNATION 2.3. CE MARK COMPLIANCEOPTICAL BENCH 3. GETTING STARTED3.1. UNPACKING THE M300EU 3.2. M300EU ANALYZER LAYOUTFigure 3-2 M300EU Internal Gas Flow 3.3.1. POWER CONNECTION3.3. ELECTRICAL CONNECTIONS M300EUAnalog Output Data Type Default Settings 3.3.2. ANALOG OUTPUT ConnectionsFigure 3-3 M300EU Analog Output Connector Output # 3.3.3. CONNECTING THE STATUS OUTPUTSAnalog Output Pin Outs Table 3-3 Status Output Pin AssignmentsMESSAGE 3.4. INITIAL OPERATION OF THE M300EU3.4.1. STARTUP AND WARM UP OF THE M300EU 3.4.2. WARNING MESSAGESThe Test functions available from the front panel of the M300EU are 3.4.3. FUNCTIONAL CHECK3.5.2.1. Calibration Gases 3.5.2. BASIC CO CALIBRATION SETUP3.5. INITIAL CALIBRATION 3.5.1. PRE-CALIBRATION STABILIZATIONSTEP 3 - SET CO SPAN GAS CONCENTRATION 3.5.3. BASIC CO CALIBRATION PROCEDURESTEP 1 - SET LIMITS STEP 2 - DILUTION RATIO4. FREQENTLY ASKED QUESTIONS & GLOSSARY Addendum to M300E/EM Manual P/N 300E 5. OPTIONAL HARDWARE AND SOFTWAREOPTION DESCRIPTIONM300EU Internal Pneumatic Flow- Zero/Span Valves OPT 50A 5.1. CALIBRATION VALVE OPTIONS FOR THE M300EU5.1.1. ZERO SPAN VALVES OPT 50A TOLL-FREE FAX TEL E-MAIL api-sales@teledyne.comSAMPLE CHAMBER 5.1.2. ZERO SPAN VALVES WITH IZS OPT 51BScrubber 5.1.3. Zero Span Valves with IZS and Shutoff VALVE OPT 51B 6. M300EU OPERATING INSTRUCTIONS reference gas path 6.2. OPERATING MODESM300EU Operating Modes sample gas stream is being routed through the CO scrubber of the autoAUTO-REF RANGE = 50.000 PPM 6.2.1. AUTO-REFERENCE MODE AREFSETUP Mode FieldParameter 6.2.2. TEST FUNCTIONSUnits Table 6-2 Test Functions DefinedANALOG OUT 6.3. SETUP Æ RNGE ANALOG OUTPUT REPORTING RANGE CONFIGURATION6.3.1. PHYSICAL RANGE VERSUS ANALOG OUTPUT REPORTING RANGES 6.3.2. ANALOG OUTPUT RANGES FOR CO CONCENTRATIONAUTO DEFAULT SETTINGSSNGL DUALThis is the default reporting range mode for the analyzer To set the ranges press following keystroke sequence M300EU OPERATING INSTRUCTIONS Consult your local regulations for the STP used by your agency 6.3.6. SETUP Æ RNGE Æ UNIT SETTING THE REPORTING RANGE UNIT TYPEweight/volume units CO ppb x 1.25 = CO ug/m3 CO ppm x 1.25= CO mg/m3 Pressure STP6.4.1. ADJUSTING THE A-REF MODE CYCLE TIME To manually start an auto-reference measurement cycle, press 6.4.2. FORCING AN AUTO-REFERENCE MEASUREMENT05514 Rev A1 DESCRIPTION 6.5. SETUP Æ MORE Æ VARS INTERNAL VARIABLES VARSTable 6-3 Test Functions Defined VARIABLEDIAG AIO 6.6. SETUP Æ MORE Æ DIAG USING THE DIAGNOSTICS FUNCTIONSDIAG I/O DIAG AOUTDIAG - Analog I/O Functions 6.7. ANALOG OUTPUT CONFIGURATION6.7.1. ANALOG I/O CONFIGURATION THE ANALOG I/O CONFIGURATION SUBMENU4. Choose a Test Channel parameter to be output on the channel 6.13.5.8 SETTING DATA UPDATE RATE FOR AN ANALOG OUTPUT ANALOG OUTPUT SIGNAL TYPE AND RANGE SELECTIONTURNING THE ANALOG OUTPUT OVER-RANGE FEATURE ON/OFF 6.13.5.7 SETTING THE REPORTING RANGE SCALE FOR AN ANALOG OUTPUTAUTOMATIC ANALOG OUTPUT CALIBRATION 6.7.2. CALIBRATION OF THE ANALOG OUTPUTSMANUAL CALIBRATION OF ANALOG OUTPUT CONFIGURED FOR VOLTAGE RANGES ENABLING OR DISABLING THE AUTO-CAL FOR AN INDIVIDUAL ANALOG OUTPUTZERO 6.7.3. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4Test Channels Functions Available on the M300EU’s Analog Output TEST CHANNELTable 6-7 M300EU Hessen GAS ID List 6.7.4. HESSEN PROTOCOLDefault Hessen Status Bit Assignments STEP3 - ZERO/SPAN CALIBRATION 7. CALIBRATION PROCEDURESmanual for troubleshooting tips MANUAL CALIBRATION REQUIRED EQUIPMENT, SUPPLIES AND EXPENDABLESMANUAL CALIBRATION CHECKS WITH ZERO/SPAN VALVES 7.9.1 SECTION O2 SENSOR CALIBRATION PROCEDURE7.9.2 CO2 SENSOR CALIBRATION PROCEDURE MANUAL CALIBRATION WITH ZERO/SPAN VALVES8. EPA PROTOCOL CALIBRATION Addendum to M300E/EM Manual P/N 9.1. PERFOMING AN AUTO-REFERENCE MEASUREMENT AFTER MAINTENANCE 9. INSTRUMENT MAINTENANCEAddendum to M300E/EM Manual P/N MR RATIO Table 9-1 M300EU Test Function RecordSTABILITY CO MEASAddendum to M300E/EM Manual P/N Figure 10-1 M300EU Gas flow during Auto-Reference Measurements 10. THEORY OF OPERATION10.1. MEASUREMENT METHOD 10.1.1. AUTO-REFERENCE RATIOThis includes calibration gases such as zero air and span gas 10.2. PNEUMATIC OPERATION10.2.1. THE A-REF CARBON MONOXIDE SCRUBBER Figure 10-2 Auto-Reference Measurement CycleFigure 10-3 Semi-Permeable Membrane Drying Process 10.2.2. THE A-REF VALVE ASSEMBLY10.2.3. THE NAFION→ DRYER Auto-Reference Valve Operating StatesThe dryer consists of a bundle of 0.030 O.D. Nafion→ tubes housed within a single large tube shell. Sample gas flows within the Nafion tubes while water vapor absorbs into the tubular membrane walls and is removed. Moisture permeating the tubing is carried away by a dry purge gas within the shell that flows countercurrent to the sample gas Aout 10.3. ELECTRONIC OPERATION10.3.1. OVERVIEW Figure 10-4 M300EU Electronic Overview Block Diagram10.3.2.1. Temperature Control of the Convection Oven 10.3.2. THE RELAY PCAFigure 10-5 M300EU Heating Control Block Diagram Figure 10-6 M300EU Oven Heater Configuration Jumpers 10.3.2.2. Oven Heaters AC Power Configuration115 VAC Configuration 230 VAC Configuration10.3.3.2. Sensor Inputs 10.3.3. MOTHERBOARD10.3.2.3. Status LED’s 10.3.3.1. A to D ConversionRELAY PCA 10.3.4. POWER DISTRIBUTIONFigure 10-8 M300EU Distribution Block Diagram Page PROCEDURES 11. TROUBLESHOOTING & REPAIR11.1. GENERAL NOTES MAINTENANCE PERSONNEL ONLYWarning Messages - Indicated Failures 11.2. FAULT DIAGNOSIS WITH WARNING MESSAGESTable continued on next page Sample flow rate is Warning Messages - Indicated Failures contINDICATED FAILURES 11.2.1. FAULT DIAGNOSIS WITH TEST FUNCTIONSTable 11-2 Test Functions - Indicated Failures table continuedValues outside range indicate Test Functions - Indicated FailuresVoltage displayed should change. If not 11.2.2. RELAY BOARD STATUS LED’SFaulty Connectors/Wiring Table 11-3 Relay Board Status LED Failure Indications11.4. OTHER PERFORMANCE PROBLEMS 11.3. GAS FLOW PROBLEMS11.4.1. Unexplained DRIFT 11.5.2.1. A/D Functions 11.5. SUBSYSTEM CHECHOUT11.5.1. RELAY BOARD 11.5.2. MOTHERBOARD11.6. TECHNICAL ASSISTANCE Addendum to M300E/EM Manual P/N M300EU Addendum to M300E/EM Manual APPENDIX A - Version Specific Software DocumentationAPPENDIX A-2 M300EU Setup Variables Available Via Serial I/O APPENDIX A-4 M300EU Signal I/O DefinitionsTEST1 CLR1MSG1 Figure A-2 Primary Setup Menu Except iDAS Figure A-3 Primary Setup Menu Æ iDAS Submenu Figure A-4 Secondary Setup Menu Æ COMM and VARS Submenus Figure A-5 Secondary Setup Menu Æ Hessen Protocol Submenu Press ENTR ENTER PASSWORDAuto Cal feature is OFF or the range is set for Figure A-6 DIAG MenuAPPENDIX A-2 Setup Variables For Serial I/O, Revision K.6 APPENDIX A-2 Setup Variables For Serial I/O, Revision K.6Table continues CO2FILTPCT O2FILTPCT A-10STDO2CELLTEMP A-11 APPENDIX A-3 Warnings and Test Functions, Revision K.6NAME MESSAGE TEXTThese gases not applicable to the M300EU Table A-5A-12 These gases not applicable to the M300EUNumber Signal NameBit or Channel DescriptionOVTMPW Internal box temperature #2/oven warning APPENDIX A-5 M300EU iDAS Parameters Revision K.6A-14 O2OFST A-15CO2OFS O2SLPEA-16 APPENDIX A-6 Terminal Command Designators, Revision K.605925 Rev A APPENDIX B - M300EU SPARE PARTS & EXPENDABLESAPPENDIX B - M300EU SPARE PARTS & EXPENDABLES European standard ENM300EU Addendum to M300E/EM Manual