M300EU Addendum to M300E/EM Manual | APPENDIX | |||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| SETUP VARIABLE |
|
| NUMERIC |
|
|
| DEFAULT |
|
|
| VALUE RANGE |
|
| DESCRIPTION |
|
|
|
| UNITS |
|
|
| VALUE |
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| OVEN_PROP |
| 1/ºC | 0.5 |
|
|
|
| Internal box temperature #2/oven PID | |||||||
|
|
|
|
|
| proportional coefficient. Proportional band is the | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| reciprocal of this setting. | |
| OVEN_INTEG |
| — | 0.02 |
|
|
|
| Internal box temperature #2/oven PID integral | |||||||
|
|
|
|
|
| coefficient. | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| OVEN_DERIV |
| — | 0 |
|
|
|
| Internal box temperature #2/oven PID derivative | |||||||
|
|
|
|
|
| coefficient. | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
| 42 |
|
|
|
|
|
| Internal box temperature #2/oven set point and | |||
| OVEN_SET |
| ºC |
|
|
|
|
|
|
| ||||||
|
|
|
| Warnings: |
|
| ||||||||||
|
|
|
|
|
| warning limits. | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
| ||||||
| BENCH_CYCLE |
| Seconds | 15 |
|
|
|
| Optical bench temperature control cycle period. | |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| BENCH_PROP |
|
|
|
|
|
|
|
|
|
|
|
| Optical bench temperature PID proportional | ||
|
| 1/ºC | 1.5 |
|
|
|
| coefficient. Proportional band is the reciprocal of | ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| this setting. | |
| BENCH_INTEG |
| — | 10.5 |
|
|
|
| Optical bench temperature PID integral | |||||||
|
|
|
|
|
| coefficient. | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| BENCH_DERIV |
| — | 0 |
|
|
|
| Optical bench temperature PID derivative | |||||||
|
|
|
|
|
| coefficient. | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| WHEEL_CYCLE |
| Seconds | 8 |
|
|
| Wheel temperature control cycle period. | ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| WHEEL_PROP |
| 1/ºC | 0.3 |
|
|
|
| Wheel temperature PID proportional coefficient. | |||||||
|
|
|
|
|
| Proportional band is the reciprocal of this setting. | ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
| WHEEL_INTEG |
| — | 0.06 |
|
|
|
| Wheel temperature PID integral coefficient. | |||||||
|
|
|
|
|
|
|
|
|
|
| ||||||
| WHEEL_DERIV |
| — | 0 |
|
|
|
| Wheel temperature PID derivative coefficient. | |||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Table
Because the CO2 and O2 sensor options are not available on the M300EU, the following variables listed in Appendix
SETUP VARIABLE
NUMERIC |
UNITS |
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
Medium Access Level Setup Variables (929 password)
CO2_DWELL | Seconds | 0.1 | Dwell time before taking each sample. | ||
|
|
|
|
| |
CO2_FILT_ADAPT | — | ON | ON, OFF | ON enables CO2 adaptive filter; OFF disables | |
it. | |||||
|
|
|
| ||
CO2_FILT_SIZE | Samples | 48 | CO2 moving average filter size. | ||
|
|
|
|
| |
CO2_FILT_ASIZE | Samples | 12 | CO2 moving average filter size in adaptive | ||
mode. | |||||
|
|
|
| ||
CO2_FILT_DELTA | % | 2 | Absolute CO2 conc. change to trigger adaptive | ||
filter. | |||||
|
|
|
| ||
CO2_FILT_PCT | % | 10 | Percent CO2 conc. change to trigger adaptive | ||
filter. | |||||
|
|
|
| ||
CO2_FILT_DELAY | Seconds | 90 | Delay before leaving CO2 adaptive filter mode. | ||
CO2_DIL_FACTOR | — | 1 | Dilution factor for CO2. Used only if is dilution | ||
enabled with FACTORY_OPT variable. | |||||
|
|
|
| ||
O2_DWELL | Seconds | 1 | Dwell time before taking each sample. | ||
|
|
|
|
| |
O2_FILT_ADAPT | — | ON | ON, OFF | ON enables O2 adaptive filter; OFF disables it. | |
O2_FILT_SIZE | Samples | 60 | O2 moving average filter size in normal mode. | ||
|
|
|
|
| |
O2_FILT_ASIZE | Samples | 10 | O2 moving average filter size in adaptive | ||
mode. | |||||
|
|
|
| ||
O2_FILT_DELTA | % | 2 | Absolute change in O2 concentration to | ||
shorten filter. | |||||
|
|
|
| ||
Table continues |
|
|
|
|
05924 Rev K.6 |