Emerson MSM400 manual Clear Liquor Press ENT To set 23.1dB 26.2dB, Autocal Span Setting Procedure

Page 20

The display will show the following.

 

 

 

 

SET

ZERO

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In clear

liquor

0

 

please wait

 

 

 

 

 

 

 

 

After a few seconds the display will show the following.

 

 

SET ZERO

0

 

 

 

 

 

In clear

liquor

0

 

Press ENT

to set

 

 

23.1dB

26.2dB

 

 

 

 

 

 

The two dB values at the bottom of the screen are the zero values for the operating frequencies of the sensor (1MHz and 3.3MHz).

Once the ENT key is pressed the zero is set. All zero reference data is now saved together with the

date of zero calibration. The screen now gives the option to re-do the zero setting or return to the AUTOCAL menu.

 

SET

ZERO

0

 

Zero is

now set

0

ESC to finish

ENT to re-do

6.4AUTOCAL - SPAN SETTING PROCEDURE

The span setting measures the signal received from a representative sample of sludge. Later when the sludge is analysed and the results entered in the control unit the system automatically calculates the correct calibration factor (SLOPE).

If suitable slurry is not available an approximate calibration can be done using slurry types chosen from a list. This proceedure is detailed in Sectiono 6.10.

At least one span measurement must be taken. However, for better accuracy it is possible to take up to three span measurements and the MSM400 will calculate the average value.

Return to the AUTOCAL menu and highlight the SETSPAN option.

 

 

 

CALIBRATION

0

 

 

 

 

 

 

 

SETZERO Max %

0

 

SETSPAN

 

 

 

LAB VALS

 

Press ENT. The following is shown on the display

 

 

SET

SPAN

0

 

 

 

 

In sample 1

0

Press

 

 

 

 

 

 

Now allow the normal or thickest (thickest is best) sludge which is to be metered, to cover the MSM400 sensor.

Press the key the display will show.

 

SET SPAN

0

 

In sample 1

0

please wait

 

IP258

20

Image 20
Contents Mobrey Contents Safety Precautions ENT Sensor Connection IVE Neutral Earth MainsDisplay Product Introduction MSM400 Slurry Monitoring System Product DescriptionSlurry Characteristics Ultrasonic Attenuation versus Suspended solidsSuspended Sensor Type MSM433 Mobrey Pipe Section SensorsSensor Type Numbering System Keypad Operation Control UnitMSM400 Displays and Keypad EMC SPECIFICATIONS--MSM400Hart Preliminary Checks Pipe Section InstallationSuspended Sensor Installation Sensor Cables Control UnitTerminal Label Function Hart Connections and Jumper Settings Sensor ConnectionsRelays Current OutputMains Supply Safety PrecautionsInitial Power UP Main Menu Toggle RUN Calibration Setup Toggle RUN Calibration Setup MonitorToggle RUN Zero Setting Initial Zero Diagnostic ParametersEX-FACTORY System Features Calibration Methods Auto CAL Auto CAL Zero Setting Procedure Rechecking ZeroAutocal Span Setting Procedure Clear Liquor Press ENT To set 23.1dB 26.2dBSample Press ENT to set 33.6dB 44.2dB Auto CAL LAB ValuesZero Setting Procedure Maximum % SolidsSpan CALIBRATION/GRADIENT Methods Calibration Method 1-SLURRY TypeAttenuation VALUE-METHOD Start on Stop on Stop if Do not start if Duty Mode Desludge ModeNo. of operation Set Default value Max value Hrs & mins INPUT↓ Output Start Time #1 hm 730 P254 Outputs Current Output Relay OperationAlarm Middle Display←−−−−−−− Lower Display Display option parameter number DescriptionSystem Test Current Output SettingsEngineering Frequency SET Hart Smart Communications SensorsPar No Appendix A1Main Menu SUB Menu Parameter Description Alarm Report DiagnosticsLogging EngineeringMin Max Default Value Parameter DescriptionParameter Description Min Max Default Value MSM400 MonitFrom list 3.6/21mA/hold last reading Min Max Default Value Ex-Factory Manufacturer’s Code Appendix A3 Hart and Psion Operating Instructions Hart ConsiderationsCalibration adjustments at operating conditions Smart Communication with the Mobrey MSM400Further customisation using the Smart HHC Hand Held Communicator Assembly Instructions Handheld Communicator Mobrey CKFig DII. Loop diagram Fig D1 MOBREY-CK* HHC assembly How to connect the Smart Communicator Hand Held Communicator Operation Mobrey Find Save Diary Calc Time Notes World Alarm How to drive a Psion based Smart CommunicatorTAG Tank Help D7.0 Introduction and Function menuD7.1 Monitor/Display Parameters D D7.2 Program Parameters PDefault Register Mode KEYWorking Register Safe RegisterHand Held Communicator Registers Select Data Source 10.0ERROR Messages on the HHC 10.1Alarm and Error MessagesPrintout or PC transfer of MSM400 programme data Display of Parameter Data10.2Invalid Data Entry Password not Open Instrument in Multidrop Mode10.3Communication Errors Invalid Action not the Same InstrumentD0 D8 Current Output ** Set current Trim-maximum Trim-maximum 11.0Current Loop Checks and TrimmingCurrent and USE ↑ & ↓ To Trim the Output 12.0Unknown InstrumentD13.0 Smart Interfaces Compatibility D13.1 Introduction 13.3Use of multiple Smart CommunicatorsAbcdef