Emerson MSM400 manual Sample Press ENT to set 33.6dB 44.2dB, Auto CAL LAB Values

Page 21

After a few secnds the display will show the following.

 

 

 

SET SPAN

0

 

 

 

 

 

 

In sample 1

0

Press ENT to set 33.6dB 44.2dB

The bottom line of the display shows the attenuation caused by the sludge. During a desludge these readings will fluctuate due to random variations in the sludge density. When the readings are reasonably stable press ENT at the same time that a sample is taken. This will store the reading from the sensor.

 

SET

SPAN

0

 

Span1 is

now set

0

ESC to finish

for Span2

Press the down arrow key to access the sample 2 and sample 3 screens (if required).

Repeat the above procedure for samples 2 and 3. The span is now set and ready for input of laboratory result (see later). When the sample procedure is complete press ESC until the normal display shows.

For best accuracy the three samples should be taken over a period of approximately 2 to 3 minutes. The samples should now sent for laboratory analysis to establish the actual suspended solids content.

When the laboratory analysis results are available, the span calibration can be completed by entering the solids content (in % Solids) into “Lab Value 1” (P150).

If required, an estimated value can be used until the laboratory analysis results are available.

Note:

Each time SPAN 1 is set the other two SPAN and LAB values are cleared and are not used

 

until new Span2 and Span3 calibrations are carried out. The MSM400 calculates the

 

average of all the valid span calibrations (a valid span calibration is one which has both a

 

span ‘N’ and lab val ‘N’ value). The calibrations must be carried out in numerical order

 

i.e. Span 1, then Span 2 and then Span 3. These values should be carried over a short

 

period of time i.e. 2 to 3 minutes.

6.5AUTO CAL LAB VALUES

To complete the AUTO CAL the samples must now be analysed and the results entered into the lab value menu. Access the LAB VALS menu by highlighting this option in the AUTOCAL menu and pressing ENT.

 

LAB VAL

0

 

LabVal 1

0

LabVal 2

LabVal 3

Highlight the number of the LabVal to be entered and press ENT

 

LAB

VAL

0

 

LabVal 1

%

0

0.00

P150

 

Highlight the digit to be edited by pressing left or right arrows. To change the LabVal use the up and down arrows until the required value is showing, then use left or right arrow to move on to next digit. The units of this parameter are always % suspended solids. When the correct value is displayed press ENT to store the value. Repeat this step to input LabVal 2 and LabVal 3 if required (please note that better accuracy is achieved if all three samples and corresponding LabVals are entered). Press ESC until the normal display shows.

It is important to note that until the LabVal (1,2 or 3) are entered the system will use the default value of attenuation for the calibration.

IP258

21

Image 21
Contents Mobrey Contents Safety Precautions Sensor Connection IVE Neutral Earth Mains DisplayENT Product Description Product Introduction MSM400 Slurry Monitoring SystemUltrasonic Attenuation versus Suspended solids Slurry CharacteristicsMobrey Pipe Section Sensors Suspended Sensor Type MSM433Sensor Type Numbering System Control Unit MSM400 Displays and KeypadKeypad Operation SPECIFICATIONS--MSM400 HartEMC Pipe Section Installation Suspended Sensor InstallationPreliminary Checks Control Unit Sensor CablesTerminal Label Function Sensor Connections Hart Connections and Jumper SettingsRelays Current OutputSafety Precautions Initial Power UPMains Supply Toggle RUN Calibration Setup Monitor Main Menu Toggle RUN Calibration SetupToggle RUN Diagnostic Parameters EX-FACTORY System Features Zero Setting Initial Zero Rechecking Zero Calibration Methods Auto CAL Auto CAL Zero Setting ProcedureClear Liquor Press ENT To set 23.1dB 26.2dB Autocal Span Setting ProcedureAuto CAL LAB Values Sample Press ENT to set 33.6dB 44.2dBMaximum % Solids Zero Setting ProcedureSpan CALIBRATION/GRADIENT Methods Calibration Method 1-SLURRY TypeAttenuation VALUE-METHOD Duty Mode Desludge Mode No. of operation Set Default value Max value Hrs & minsStart on Stop on Stop if Do not start if INPUT↓ Output Start Time #1 hm 730 P254 Relay Operation Outputs Current OutputMiddle Display Alarm←−−−−−−− Lower Display Display option parameter number DescriptionSettings Engineering Frequency SETSystem Test Current Output Sensors Hart Smart CommunicationsAppendix A1 Main Menu SUB Menu Parameter DescriptionPar No Diagnostics Alarm ReportLogging EngineeringParameter Description Min Max Default ValueMSM400 Monit Parameter Description Min Max Default ValueFrom list 3.6/21mA/hold last reading Min Max Default Value Ex-Factory Manufacturer’s Code Hart Considerations Appendix A3 Hart and Psion Operating InstructionsSmart Communication with the Mobrey MSM400 Calibration adjustments at operating conditionsFurther customisation using the Smart HHC Handheld Communicator Mobrey CK Hand Held Communicator Assembly InstructionsFig DII. Loop diagram Fig D1 MOBREY-CK* HHC assembly How to connect the Smart Communicator Hand Held Communicator Operation How to drive a Psion based Smart Communicator Mobrey Find Save Diary Calc Time Notes World AlarmTAG Tank D7.0 Introduction and Function menu HelpD7.2 Program Parameters P D7.1 Monitor/Display Parameters DMode KEY Default RegisterWorking Register Safe RegisterHand Held Communicator Registers Select Data Source 10.1Alarm and Error Messages 10.0ERROR Messages on the HHCPrintout or PC transfer of MSM400 programme data Display of Parameter Data10.2Invalid Data Entry Instrument in Multidrop Mode Password not Open10.3Communication Errors Invalid Action not the Same InstrumentD0 D8 11.0Current Loop Checks and Trimming Current Output ** Set current Trim-maximum Trim-maximum12.0Unknown Instrument Current and USE ↑ & ↓ To Trim the Output13.3Use of multiple Smart Communicators D13.0 Smart Interfaces Compatibility D13.1 IntroductionAbcdef