Emerson MSM400 manual Manufacturer’s Code

Page 37

Parameter

Description

 

 

 

Min.

Max.

Default

No.

 

 

 

 

value

value

Ex-Factory

D750

Model Code

Hart parameter

-

-

-

D751

Serial Number - Control Unit

-

-

-

 

Store as Final Assembly Number (FAN)

 

 

 

D752

Hardware Revision

 

 

-

-

-

 

Factory-set

 

 

 

 

 

 

D753

Software Revision

 

 

-

-

-

 

Factory-set, embedded in software

 

 

 

D760

Manufacturer’s Code

 

 

-

-

-

 

Factory set not atlerable

 

 

 

 

D761

Unique ID

 

 

 

-

-

-

 

Hart parameter

 

 

 

 

 

D762

Universal Command Revision

-

-

-

 

Hart parameter

 

 

 

 

 

D763

Transmitter Specific Command Revision

-

-

-

 

Hart parameter

 

 

 

 

 

D764

Preamble Bytes

 

 

-

-

-

 

Hart parameter

 

 

 

 

 

D765

Flags Hart parameter

 

-

-

-

D800

PV

 

 

 

-

-

-

D801

Sludge Density

 

 

-

-

-

D805

% Current Output

 

 

-

-

-

D806

Current output

Current is proportional to PV

-

-

-

D810

Totaliser Value

 

 

-

-

-

D820

Relay Status

 

 

 

-

-

-

D821

Relay 1 Run-Time

Resettable - Hours & Minutes

-

-

-

D822

Relay 2 Run-Time

Resettable - Hours & Minutes

-

-

-

D830

Alarm report

e.g. Sensor Fault, ….

-

-

-

D835

Digital input status

 

 

-

-

-

D841

Current input

 

 

 

 

 

 

 

Value of current input (mA)

 

-

-

-

D844

Temperature

 

 

 

 

 

 

 

The temperature oC within the control unit

-

-

-

D850

Attenuation @ 1MHz Attenuation

-

-

-

D851

Attenuation @ 3.3MHz

Attenuation

-

-

-

D852

Attenuation unsmoothed

 

-

-

-

D853

Signal level (Raw A/D bits)

 

 

 

 

 

Raw RF A/D output

 

 

-

-

-

D854

Span Average @ 1MHz (dB)

 

 

 

 

 

Diagnostic parameter that averages the span calibrations

-

-

-

D855

Span Average @ 3.3MHz

 

 

 

 

 

Diagnostic parameter that averages the span calibrations

-

-

-

D858

Slope 1

 

 

 

 

 

 

 

The value of the attenuation in dB/% @ 1MHZ

 

 

 

 

determined by the control unit.

-

-

-

D859

Slope 3

 

 

 

 

 

 

 

The value of the attenuation in dB/% @ 3.3MHZ

 

 

 

 

determined by the control unit.

-

-

-

D860

Frequency of operation

 

 

 

 

 

Diagnostic of frequency(ies) being used (MHz)

-

-

-

D861

Maximum density

 

 

 

 

 

 

The maximum density that can be measured.

 

 

 

 

This value is calculated from zero and span data.

-

-

-

IP258

37

Image 37
Contents Mobrey Contents Safety Precautions Display Sensor Connection IVE Neutral Earth MainsENT Product Description Product Introduction MSM400 Slurry Monitoring SystemUltrasonic Attenuation versus Suspended solids Slurry CharacteristicsMobrey Pipe Section Sensors Suspended Sensor Type MSM433Sensor Type Numbering System MSM400 Displays and Keypad Control UnitKeypad Operation Hart SPECIFICATIONS--MSM400EMC Suspended Sensor Installation Pipe Section InstallationPreliminary Checks Control Unit Sensor CablesTerminal Label Function Sensor Connections Hart Connections and Jumper SettingsRelays Current OutputInitial Power UP Safety PrecautionsMains Supply Toggle RUN Calibration Setup Monitor Main Menu Toggle RUN Calibration SetupToggle RUN EX-FACTORY System Features Diagnostic ParametersZero Setting Initial Zero Rechecking Zero Calibration Methods Auto CAL Auto CAL Zero Setting ProcedureClear Liquor Press ENT To set 23.1dB 26.2dB Autocal Span Setting ProcedureAuto CAL LAB Values Sample Press ENT to set 33.6dB 44.2dBMaximum % Solids Zero Setting ProcedureSpan CALIBRATION/GRADIENT Methods Calibration Method 1-SLURRY TypeAttenuation VALUE-METHOD No. of operation Set Default value Max value Hrs & mins Duty Mode Desludge ModeStart on Stop on Stop if Do not start if INPUT↓ Output Start Time #1 hm 730 P254 Relay Operation Outputs Current OutputMiddle Display Alarm←−−−−−−− Lower Display Display option parameter number DescriptionEngineering Frequency SET SettingsSystem Test Current Output Sensors Hart Smart CommunicationsMain Menu SUB Menu Parameter Description Appendix A1Par No Diagnostics Alarm ReportLogging EngineeringParameter Description Min Max Default ValueMSM400 Monit Parameter Description Min Max Default ValueFrom list 3.6/21mA/hold last reading Min Max Default Value Ex-Factory Manufacturer’s Code Hart Considerations Appendix A3 Hart and Psion Operating InstructionsSmart Communication with the Mobrey MSM400 Calibration adjustments at operating conditionsFurther customisation using the Smart HHC Handheld Communicator Mobrey CK Hand Held Communicator Assembly InstructionsFig DII. Loop diagram Fig D1 MOBREY-CK* HHC assembly How to connect the Smart Communicator Hand Held Communicator Operation How to drive a Psion based Smart Communicator Mobrey Find Save Diary Calc Time Notes World AlarmTAG Tank D7.0 Introduction and Function menu HelpD7.2 Program Parameters P D7.1 Monitor/Display Parameters DMode KEY Default RegisterWorking Register Safe RegisterHand Held Communicator Registers Select Data Source 10.1Alarm and Error Messages 10.0ERROR Messages on the HHCPrintout or PC transfer of MSM400 programme data Display of Parameter Data10.2Invalid Data Entry Instrument in Multidrop Mode Password not Open10.3Communication Errors Invalid Action not the Same InstrumentD0 D8 11.0Current Loop Checks and Trimming Current Output ** Set current Trim-maximum Trim-maximum12.0Unknown Instrument Current and USE ↑ & ↓ To Trim the Output13.3Use of multiple Smart Communicators D13.0 Smart Interfaces Compatibility D13.1 IntroductionAbcdef