Emerson MSM400 manual From list 3.6/21mA/hold last reading

Page 35

Parameter

Description

No.

 

P400

Lower range value

 

The mA set point for minimum PV in selected units

P401

Upper range value

 

The mA set point for maximum PV in selected units

P402

Alarm action

 

The is the value that the current will drive to on alarm

 

action. The alarm action is programable and is selected

 

from list (3.6/21mA/hold last reading).

P403

0-20mA/4-20mA

 

The mA output is selectable by user,

 

either 0 or 4 to 20mA

P410

Relay 1 Mode

 

Select Relay function from a list (Off, Set Point,Desludge

 

Alarm, Totaliser, Fault)

P411

Relay 1 On Point

 

The value of the variable (as set in relay mode parameter)

 

at which relay 1 turns on. Programmable by user

P412

Relay 1 Off Point

 

The value of the variable (as set in relay mode parameter)

 

at which relay 1 turns off.

P413

Relay 1 Min On Time

 

A programmable parameter for the minimum on time

 

of relay 1. This is to overcome slugs of clear liquor at

 

the start of desludge cycles

P414

Relay1 Max On Time

 

A programmable parameter for the maximum on time

 

of relay 1. This is to ensure, for example, a pump does

 

not continue to run due to a faulty signal from sensor.

P420

Relay 2 Mode

 

Select Relay function from a list (Off, Set Point,Desludge

 

Alarm, Totaliser, Fault)

P421

Relay 2 On Point

 

The value of the variable (as set in relay mode parameter)

 

at which relay 2 turns on. Programmable by user

P422

Relay 2 Off Point

 

The value of the variable (as set in relay mode parameter)

 

at which relay 2 turns off. Programmable by user

P423

Relay 2 Min On Time

 

A programmable parameter for the minimum on time

 

of relay 2. This is to overcome slugs of clear liquor at

 

the start of desludge cycles

P424

Relay 2 Max On Time

 

A programmable parameter for the maximum on time

 

of relay off. This is to ensure, for example, a pump does

 

not continue to run due to a faulty signal from sensor.

P415

Relay 1 Min Off Time

P425

Relay 2 Min Off Time

Min.

Max.

Default

value

value

Ex-Factory

-

-

0

-

-

auto

1

3

hold

1

2

4-20mA

1

5

set point

-

-

0

-

-

0

0:00

999.59

000:00

0:00

999.59

000:00

1

5

set point

-

-

0

-

-

0

0:00

999.59

000:00

0:00

999.59

000:00

-

-

-

-

-

-

IP258

35

Image 35
Contents Mobrey Contents Safety Precautions ENT Sensor Connection IVE Neutral Earth MainsDisplay Product Description Product Introduction MSM400 Slurry Monitoring SystemUltrasonic Attenuation versus Suspended solids Slurry CharacteristicsMobrey Pipe Section Sensors Suspended Sensor Type MSM433Sensor Type Numbering System Keypad Operation Control UnitMSM400 Displays and Keypad EMC SPECIFICATIONS--MSM400Hart Preliminary Checks Pipe Section InstallationSuspended Sensor Installation Control Unit Sensor CablesTerminal Label Function Current Output Hart Connections and Jumper SettingsSensor Connections RelaysMains Supply Safety PrecautionsInitial Power UP Toggle RUN Calibration Setup Monitor Main Menu Toggle RUN Calibration SetupToggle RUN Zero Setting Initial Zero Diagnostic ParametersEX-FACTORY System Features Rechecking Zero Calibration Methods Auto CAL Auto CAL Zero Setting ProcedureClear Liquor Press ENT To set 23.1dB 26.2dB Autocal Span Setting ProcedureAuto CAL LAB Values Sample Press ENT to set 33.6dB 44.2dBCalibration Method 1-SLURRY Type Zero Setting ProcedureMaximum % Solids Span CALIBRATION/GRADIENT MethodsAttenuation VALUE-METHOD Start on Stop on Stop if Do not start if Duty Mode Desludge ModeNo. of operation Set Default value Max value Hrs & mins INPUT↓ Output Start Time #1 hm 730 P254 Relay Operation Outputs Current OutputDisplay option parameter number Description AlarmMiddle Display ←−−−−−−− Lower DisplaySystem Test Current Output SettingsEngineering Frequency SET Sensors Hart Smart CommunicationsPar No Appendix A1Main Menu SUB Menu Parameter Description Engineering Alarm ReportDiagnostics LoggingParameter Description Min Max Default ValueMSM400 Monit Parameter Description Min Max Default ValueFrom list 3.6/21mA/hold last reading Min Max Default Value Ex-Factory Manufacturer’s Code Hart Considerations Appendix A3 Hart and Psion Operating InstructionsSmart Communication with the Mobrey MSM400 Calibration adjustments at operating conditionsFurther customisation using the Smart HHC Handheld Communicator Mobrey CK Hand Held Communicator Assembly InstructionsFig DII. Loop diagram Fig D1 MOBREY-CK* HHC assembly How to connect the Smart Communicator Hand Held Communicator Operation How to drive a Psion based Smart Communicator Mobrey Find Save Diary Calc Time Notes World AlarmTAG Tank D7.0 Introduction and Function menu HelpD7.2 Program Parameters P D7.1 Monitor/Display Parameters DSafe Register Default RegisterMode KEY Working RegisterHand Held Communicator Registers Select Data Source Display of Parameter Data 10.0ERROR Messages on the HHC10.1Alarm and Error Messages Printout or PC transfer of MSM400 programme data10.2Invalid Data Entry Invalid Action not the Same Instrument Password not OpenInstrument in Multidrop Mode 10.3Communication ErrorsD0 D8 11.0Current Loop Checks and Trimming Current Output ** Set current Trim-maximum Trim-maximum12.0Unknown Instrument Current and USE ↑ & ↓ To Trim the Output13.3Use of multiple Smart Communicators D13.0 Smart Interfaces Compatibility D13.1 IntroductionAbcdef