APPENDIX A2: FULL LIST OF FUNCTIONS

Parameter list and description

P=Parameter,

D=Diagnostic Display Parameter

Parameter

Description

No.

 

P100

Sensor Gap in mm.

 

This is a user entered parameter (optional).

 

This value is ignored when an alternative span

 

calibration is caried out (i.e. method two or three).

P101

Sludge type

 

The user can select from a list of sludge type to set

 

up the span for the calibration procedure (optional).

 

Each sludge type is associated with an attenuation

 

factor for each frequency. This value is ignored when

 

an alternative span calibration is caried out

 

(i.e. method two or three).

P102

dB Factor @ 1MHz

 

A user alterable parameter. This is the attenuation

 

number in dB/% at 1MHz. This is used for the span

 

calibration (method three).

P103

dB Factor @ 3.3MHz

 

A user alterable parameter. This is the attenuation

 

number in dB/% at 3.3MHz.This is used for the span

 

calibration (method three).

P120

Zero Ref @ 1MHz

 

This is the value in dBs (@ 1MHz) that is stored in

 

the control unit when a zero calibration is carried out.

 

This value is over written each time a zero calibration

 

is carried out.

P121

Zero Ref @ 3.3MHz

 

This is the value in dBs (@ 3.3MHz) that is stored in

 

the control unit when a zero calibration is carried out.

 

This value is over written each time a zero calibration

 

is carried out.

P122

Date of Zero Ref

 

This is automatically stored by the control unit when

 

the zero calibration is done. The actual date can be

 

set-up in the ENGINEER menu (P730)

P123

Initial Zero Ref @ 1MHz dB’s

 

This is the value in dBs (@ 1MHz) that is stored in

 

the control unit when the initial zero calibration is

 

carried out.This value is compared with subsequent

 

zero calibrations.0 must be entered to reset or clear initial

 

zero value

P124

Initial Zero Ref @ 3.3MHz dB’s

 

This is the value in dBs (@ 3.3MHz) that is stored in

 

the control unit when the initial zero calibration is

carried out.This value is compared with subsequent

zero calibrations.0 must be entered to reset or clear initial

Min

Max

Default

value

value

Ex-Factory

0

1000

0.0

-

-

none

0

-

0.000

0

-

0.000

-

-

16

-

-

16

-

-

--/--/----

-

-

0

IP258

33

Image 33
Contents Mobrey Contents Safety Precautions Sensor Connection IVE Neutral Earth Mains DisplayENT Product Description Product Introduction MSM400 Slurry Monitoring SystemUltrasonic Attenuation versus Suspended solids Slurry CharacteristicsMobrey Pipe Section Sensors Suspended Sensor Type MSM433Sensor Type Numbering System Control Unit MSM400 Displays and KeypadKeypad Operation SPECIFICATIONS--MSM400 HartEMC Pipe Section Installation Suspended Sensor InstallationPreliminary Checks Control Unit Sensor CablesTerminal Label Function Sensor Connections Hart Connections and Jumper SettingsRelays Current OutputSafety Precautions Initial Power UPMains Supply Toggle RUN Calibration Setup Monitor Main Menu Toggle RUN Calibration SetupToggle RUN Diagnostic Parameters EX-FACTORY System FeaturesZero Setting Initial Zero Rechecking Zero Calibration Methods Auto CAL Auto CAL Zero Setting ProcedureClear Liquor Press ENT To set 23.1dB 26.2dB Autocal Span Setting ProcedureAuto CAL LAB Values Sample Press ENT to set 33.6dB 44.2dBMaximum % Solids Zero Setting ProcedureSpan CALIBRATION/GRADIENT Methods Calibration Method 1-SLURRY TypeAttenuation VALUE-METHOD Duty Mode Desludge Mode No. of operation Set Default value Max value Hrs & minsStart on Stop on Stop if Do not start if INPUT↓ Output Start Time #1 hm 730 P254 Relay Operation Outputs Current OutputMiddle Display Alarm←−−−−−−− Lower Display Display option parameter number DescriptionSettings Engineering Frequency SETSystem Test Current Output Sensors Hart Smart CommunicationsAppendix A1 Main Menu SUB Menu Parameter DescriptionPar No Diagnostics Alarm ReportLogging EngineeringParameter Description Min Max Default ValueMSM400 Monit Parameter Description Min Max Default ValueFrom list 3.6/21mA/hold last reading Min Max Default Value Ex-Factory Manufacturer’s Code Hart Considerations Appendix A3 Hart and Psion Operating InstructionsSmart Communication with the Mobrey MSM400 Calibration adjustments at operating conditionsFurther customisation using the Smart HHC Handheld Communicator Mobrey CK Hand Held Communicator Assembly InstructionsFig DII. Loop diagram Fig D1 MOBREY-CK* HHC assembly How to connect the Smart Communicator Hand Held Communicator Operation How to drive a Psion based Smart Communicator Mobrey Find Save Diary Calc Time Notes World AlarmTAG Tank D7.0 Introduction and Function menu HelpD7.2 Program Parameters P D7.1 Monitor/Display Parameters DMode KEY Default RegisterWorking Register Safe RegisterHand Held Communicator Registers Select Data Source 10.1Alarm and Error Messages 10.0ERROR Messages on the HHCPrintout or PC transfer of MSM400 programme data Display of Parameter Data10.2Invalid Data Entry Instrument in Multidrop Mode Password not Open10.3Communication Errors Invalid Action not the Same InstrumentD0 D8 11.0Current Loop Checks and Trimming Current Output ** Set current Trim-maximum Trim-maximum12.0Unknown Instrument Current and USE ↑ & ↓ To Trim the Output13.3Use of multiple Smart Communicators D13.0 Smart Interfaces Compatibility D13.1 IntroductionAbcdef