External equipment intended for connection to signal input, signal output, or other connectors, must comply with the relevant IEC standard (EN/IEC 60601-1 series for medical electrical equipment). In addition, all such combinations (systems) must comply with the standard IEC 60601-1-1, Safety requirements for medical electrical systems.

Equipment not complying to IEC 60601 must be kept outside the patient environment, as defined in the standard as at least 1.5 meters from the patient or the patient support.

Any person who connects external equipment to signal input, signal output, or other connectors has formed a system and is therefore responsible for the system to comply with the requirements of IEC 60601-1-1. If in doubt, speak

with a qualified technician.

Safety tips

Never open the display case, even when the power is off. Dangerous voltage inside may cause electric shock or death.

To avoid damage to the display, use the grounded power supply and video cable supplied by Planar, or use certified replacements.

Be sure the display is electrically grounded. You must connect the third grounding pin on the US power cord to a grounded outlet. The European power cord does not have a third grounding pin, but it must be plugged into a grounded outlet.

If you cannot insert the plug into the outlet you plan to use, have a licensed electrician replace the outlet with a properly grounded outlet. If the power cord connects directly into the computer, make sure the computer is grounded.

Keep the display dry if it is part of a surgical system. The display lacks protection against liquids or spills.

In locations where 240V outlets are used, connect the Dome EX display only to a center-tapped, 240V, single-phase supply (only for Canada and the United States).

GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN EQUIPMENT IS CONNECTED TO AN EQUIVALENT RECEPTABLE MARKED “HOSPITAL ONLY” OR “HOSPITAL GRADE.”

vi Dome EX Display

Image 6
Contents Dome EX Display User’s GuidePlanar Systems, Inc. All rights reserved Contents Product Information Symbol explanations Intended useSafety precautions Safety tipsUnpacking and handling tips Preventing fire and injuryCleaning the display Shipping/storing the display Disposal informationDVI cable About the Display System Requirements Display Components Tilt range Height range Orientation modeDisplay Positions Viewing angleStand cable cover Desk Stand FeaturesStand height lock USB hubInstalling the Display Set the DIP Switch on the Dome DX2 Board TipsInstall the Display Controller To install the display controllerConnect the Video and Power Cables To connect the cablesTo enable the display Install the Dome DX Display DriverTo install the driver To change the display propertiesTo set brightness To set the resolution and paletteAuto-orientation To set driver optionsInstall the AX3 Display Driver To configure the displayTo adjust the display properties To set the display mode manuallyTo use a dual-display configuration Appendix Troubleshooting Problem Possible Cause SolutionDome E2 Specification Category Characteristic Item SpecificationPower Supply ReliabilityEnvironment Dome E3 Specification ModelDome E3 Specification Dome E3 Specification EMI shielding No emission of low-level radiation Temperature Dome E3c Specification Display Mtbf 30,000 hours Backlight Dome E5 Specification Dome E5 Specification Video Modes Connector Ports Details of DIN connectorsLED Status Lights Power-up sequence information from LED LED a status information after initial power-onLED B status information after initial power-on To set the display resolution Single DesktopTo change the configuration to single desktop Restoring dual headsAuto-Installation System configuration requirementInstallation options Prompting user response during Fastinst.bat Display Driver Removal To remove the Dome DX driverTo remove the AX3 driver Power Management Grayscale Palette Options Dynamic gray palette Palette options with multiple displaysDesktop icons Nonlinear static gray paletteStatic gray palette Index Using single desktop Regulatory Compliance Canada, European Union, United StatesPage Fax + 358 9 420 EU Declaration of Conformity for Medical Application FCC Compliance Statement Japan AustralianChina KoreaStandard Warranty Standard Warranty Return ProcedureSummary Summary Limitations and Exclusions of Dome Displays Limitation of Implied WarrantiesExclusion of Damages Effect of Local LawMedicalsales@planar.com Medicalsupport@planar.com