Sigma SQM-160 manual Chapter Operation Sensor Selection

Page 18

Chapter 2

Operation

2.4Sensor Selection

The SQM-160 comes standard with two sensor inputs. Four additional sensors are available by adding a Sensor Option Card. The sensors can each be assigned to a different film, or multiple sensors can be averaged for a single film. The averaging option provides more uniform coverage of the deposition area, and provides a backup sensor capability. If one of multiple sensors assigned to a film fails, the sensor is automatically removed from rate/thickness calculations.

To assign a sensor, or sensors, to a film:

1.Press Program to enter program mode.

2.Use the Control Knob to scroll to the desired Film # (1-9).

3.Depress the Control Knob or Next to enter the film parameters menu for the selected film.

4.Press Next until SENS AVG is shown.

5.Use the Control Knob to scroll through the sensors in Display2.

6.Depress the Control Knob to toggle the sensor on/off.

Sensor status can be seen by observing the Crystal Status LEDs: If the LED is not illuminated, the crystal is disabled.

If the LED is illuminated,the crystal is enabled and receiving valid readings.

If the LED is flashing, the crystal is enabled, but is not receiving valid readings.

7.Continue selecting sensors until the Crystal Status LEDs indicate the desired setup.

8.Press Program to exit the Film Parameters Menu and return to normal mode.

2-7

Image 18
Contents SQM-160 Safety Information Safety SymbolsWarranty Information Limitation of WarrantyReturn Policy Chapter Options Table of Contents Chapter Quick StartChapter Operation Chapter MonComm SoftwareChapter Maintenance AppendixInstallation Chapter Quick Start IntroductionChapter Quick Start Front Panel ControlsChapter Quick Start Rear Panel Rear Panel ConnectionsChapter Quick Start System Connections System ComponentsChapter Quick Start Film Setup Chapter Quick Start Depositing a Film Chapter Operation Introduction Menu SelectionChapter Operation Film Parameters Setup Description Range Default Units Film Selection Film Parameters Menu Sub-MenuDisplay Chapter Operation System Parameters Setup Timebase SIM Mode FRQ Disp Rate RES Ratefilt Baudrate Etch System Parameters Menu Sub-MenuDisplay xTOOLING FMIN/MAX Freq MIN Freq MAXChapter Operation Sensor Selection Chapter Operation Sensor Min/Max Frequency Chapter Operation Sensor Tooling Chapter Operation Rate/Thickness or Frequency Display Crystal Life DisplaySimulate Mode Chapter Operation Zero ThicknessShutter Operation Chapter Options Introduction Chapter Options Four Sensor Card Analog Output CardChapter MonComm Software Introduction Run MonCommChapter MonComm Software Add and Delete Films To add a new film to the Films ListTo delete a film from the Films List To remove a film from the Download List Chapter MonComm Software Download FilmsTo add or replace a film in the Download List To download the films in the Download ListChapter MonComm Software Graph Data To Adjust the Graph’s Scope Save DataLoad Defaults Chapter MonComm Software Edit System ParametersChapter Maintenance Introduction Fuse ReplacementAppendix Material Parameters Appendix Specifications Relay I/O Connector Rear View Pins Type DescriptionAppendix O Connections Declares that the product SafetyAppendix Declaration of Conformity Product Model SQM-160 Product Options All Options