Appendix
C.I/O Connections
A
1  | 8  | 
9  | 1 5  | 
Relay I/O Connector Rear View
Relays are rated for 125VAC or 30VDC, 2A maximum. Inputs can be activated either by connecting to a switch and shorting to Ground, or they can be driven by a TTL compatible signal. These are not isolated inputs! The voltage level applied must be limited to between 0 and +5 volts with respect to Ground.
Pins | Type | Description | 
1,2  | Relay  | Crystal Fail Relay, contacts closed when all  | 
  | 
  | enabled sensors have failed.  | 
3,4  | Relay  | Timer Setpoint, contacts closed when timer  | 
  | 
  | reaches Zero from its programmed value.  | 
5,6  | Relay  | Shutter Relay, contacts closed when shutter  | 
  | 
  | is open.  | 
7,8  | Relay  | Thickness Setpoint Relay, contacts close  | 
  | 
  | when thickness value reached.  | 
9  | Input  | Zero Timer Input, grounding this pin will zero  | 
  | 
  | the timer.  | 
10  | Input  | Zero Thickness Input, grounding this pin will  | 
  | 
  | zero the thickness reading.  | 
11  | Input  | Close Shutter Input, grounding this pin will  | 
  | 
  | close the shutter.  | 
12  | Input  | Open Shutter Input, grounding this pin will  | 
  | 
  | open the shutter.  | 
13,14,15  | Ground  | 
  |