Sigma SQM-160 manual Film Selection Film Parameters Menu Sub-Menu, Display

Page 14

Chapter 2

Operation

The diagram and table below detail the parameters available in the Film Parameters Menu:

Film Selection Film Parameters Menu

Sub-Menu

Program

Film1

.

.

.

.

.

.

Film9

- DENSITY

- TOOLING

- Z-FACTOR

- FINL THK

- THK SET

- TIME SET

- SENS AVG

-1

-2

- 3

Sensors 3 to 6 show only if the

. four sensor

. option card is - 6 iinstalled.

Display

DENSITY

TOOLING

Description

Range

Default

Units

Density of the material being

0.5 – 99.99

1.00

gm/cc

deposited.

 

 

 

Overall tooling Factor for this film.

10 – 399

100

%

See the Sensor Tooling section of

 

 

 

this chapter for information on setting tooling values.

Z-FACTOR

FINL THK

Z-Factor of the material being

0.10 – 10.00

1.0

 

deposited.

 

 

 

Desired Final Thickness of

0.000 – 99.99

0.500

deposited material. Lights Final Thk

 

 

 

LED. Zero is disabled.

 

 

 

THK SET Thickness value that closes the 0.000 – 99.99 0KÅ Thickness Setpoint relay and lights

Thk SP LED. Zero is disabled.

TIME SET

SENS AVG

Elapsed time that closes the Timer

0:00 – 99:59

0

Sec

Setpoint relay and lights Time SP

 

 

 

LED. Zero is disabled.

 

 

 

Enable/disable crystals for this film.

Enabled/

Ch1

 

See the Sensor Selection section of

Disabled

Enabled

 

this chapter for more information.

 

 

 

2-3

Image 14
Contents SQM-160 Safety Information Safety SymbolsReturn Policy Warranty InformationLimitation of Warranty Chapter Options Table of Contents Chapter Quick StartChapter Operation Chapter MonComm SoftwareChapter Maintenance AppendixInstallation Chapter Quick Start IntroductionChapter Quick Start Front Panel ControlsChapter Quick Start Rear Panel Rear Panel ConnectionsChapter Quick Start System Connections System ComponentsChapter Quick Start Film Setup Chapter Quick Start Depositing a Film Chapter Operation Introduction Menu SelectionChapter Operation Film Parameters Setup Display Description Range Default UnitsFilm Selection Film Parameters Menu Sub-Menu Chapter Operation System Parameters Setup Timebase SIM Mode FRQ Disp Rate RES Ratefilt Baudrate Etch System Parameters Menu Sub-MenuDisplay xTOOLING FMIN/MAX Freq MIN Freq MAXChapter Operation Sensor Selection Chapter Operation Sensor Min/Max Frequency Chapter Operation Sensor Tooling Chapter Operation Rate/Thickness or Frequency Display Crystal Life DisplayShutter Operation Simulate ModeChapter Operation Zero Thickness Chapter Options Introduction Chapter Options Four Sensor Card Analog Output CardChapter MonComm Software Introduction Run MonCommTo delete a film from the Films List Chapter MonComm Software Add and Delete FilmsTo add a new film to the Films List To remove a film from the Download List Chapter MonComm Software Download FilmsTo add or replace a film in the Download List To download the films in the Download ListChapter MonComm Software Graph Data To Adjust the Graph’s Scope Save DataLoad Defaults Chapter MonComm Software Edit System ParametersChapter Maintenance Introduction Fuse ReplacementAppendix Material Parameters Appendix Specifications Appendix O Connections Relay I/O Connector Rear ViewPins Type Description Declares that the product SafetyAppendix Declaration of Conformity Product Model SQM-160 Product Options All Options