Sigma SQM-160 manual Chapter Operation Sensor Tooling

Page 20

Chapter 2

Operation

2.6Sensor Tooling

Sensor Tooling adjusts for the difference in deposition rate between the sensor and the substrate being coated. It is an empirically determined value that matches the sensor readings to your vacuum system.

Substrate

Substrate

Tooling

Tooling

Over 100%

Under 100%

 

The first Tooling is set in the Film Parameters Menu and is applied to the averaged Rate and Thickness for all sensors assigned to that film. Film Tooling is a film specific value.

The second Tooling Factor, xTooling, is set in the System Parameters menu. It adjusts the tooling for each individual sensor before it is averaged. xTooling for a sensor applies to all films. If the individual sensor xToolings are set properly, a sensor failure will not cause a jump in the average Rate and Thickness reading.

To adjust xTooling:

1.Press Program to enter program mode.

2.Press Prev to enter the System Parameters Menu

3.Press Next until FRQ DISP is shown, then press the Control Knob .

4.Adjust the Control Knob to set the XTLTOOL 1 value

5.Depress the Control Knob to save the value and move to XTALTOOL 2 .

6.Repeat Steps 4-5 for each of the installed sensors.

7.Press Program to exit the System Parameters Menu and return to normal mode.

2-9

Image 20
Contents SQM-160 Safety Information Safety SymbolsReturn Policy Warranty InformationLimitation of Warranty Table of Contents Chapter Quick Start Chapter OperationChapter Options Chapter MonComm SoftwareChapter Maintenance AppendixInstallation Chapter Quick Start IntroductionChapter Quick Start Front Panel ControlsChapter Quick Start Rear Panel Rear Panel ConnectionsChapter Quick Start System Connections System ComponentsChapter Quick Start Film Setup Chapter Quick Start Depositing a Film Chapter Operation Introduction Menu SelectionChapter Operation Film Parameters Setup Display Description Range Default UnitsFilm Selection Film Parameters Menu Sub-Menu Chapter Operation System Parameters Setup Timebase SIM Mode FRQ Disp Rate RES Ratefilt Baudrate Etch System Parameters Menu Sub-MenuDisplay xTOOLING FMIN/MAX Freq MIN Freq MAXChapter Operation Sensor Selection Chapter Operation Sensor Min/Max Frequency Chapter Operation Sensor Tooling Chapter Operation Rate/Thickness or Frequency Display Crystal Life DisplayShutter Operation Simulate ModeChapter Operation Zero Thickness Chapter Options Introduction Chapter Options Four Sensor Card Analog Output CardChapter MonComm Software Introduction Run MonCommTo delete a film from the Films List Chapter MonComm Software Add and Delete FilmsTo add a new film to the Films List Chapter MonComm Software Download Films To add or replace a film in the Download ListTo remove a film from the Download List To download the films in the Download ListChapter MonComm Software Graph Data To Adjust the Graph’s Scope Save DataLoad Defaults Chapter MonComm Software Edit System ParametersChapter Maintenance Introduction Fuse ReplacementAppendix Material Parameters Appendix Specifications Appendix O Connections Relay I/O Connector Rear ViewPins Type Description Safety Appendix Declaration of ConformityDeclares that the product Product Model SQM-160 Product Options All Options