VXI VM4016 user manual Inputrange

Page 31

www.vxitech.com

INPUT:RANGE

The input range command selects the input range of one or more channels. The input range may be either set for ±10 volts or ± 100 volts. It is important to note that the input offset is normalized to ± 10 volt range. The actual input offset in the 100 volt range is ten times the set value.

INPut:RANGe <range>,<channel_list>

Where <range> is 10 V 100 V.

 

Where <channel_list> is the standard

 

channel list format supporting channels 1

 

through 16.

 

 

EXAMPLES

 

INPut:RANGe 100,(@1:16)

Sets the input range for Channels 1 through

 

16 to 100 V.

INPut:RANGe? 7

Returns the input range for Channel 7 as

100

100 V.

INP:RANG 10,(@4:6)

Sets the input range for Channels 4 through

 

6 to 10 V.

INP:RANG? 5

Returns the input range for Channel 5 as

10

10 V.

VM4016 Programming

31

Image 31
Contents VM4016 Page Table of Contents VXI Technology, Inc Restricted Rights Legend WarrantyLimitation of Warranty CertificationEMC C L a R a T I O N O F C O N F O R M I T YTerms and Symbols General Safety InstructionsThis product. Product should be inspected or serviced Support Resources VXI Technology, Inc VM4016 Preface Introduction IntroductionFront Panel Layout DescriptionVM4016 Block Diagram IRQ Output VM4016 General SpecificationsCalculating System Power and Cooling Requirements InstallationPreparation for USE Setting the Logical Address Setting the Chassis Backplane JumpersFront Panel Interface Wiring Analog Comparator PIN Outs Signal PIN Number1631 Programming Examples of Scpi Commands FETCHCONDITIONED?FETCHLATCHED? FETCHRAW? Inhousepseudo Inhouseregint InhouseregintInhouseregenable Inhouseclearlatch Examples InputdebounceINPMASK? InputmaskInputmaskinterrupt INPutMASKINTerrupt boolean Where boolean is 0 OFF 1 onINPOFFS? InputoffsetINPutPOLarity NORM,@35 InputpolarityInputrange Outputpolarityexternalinterrupt Outputpolarityexternallatched Example Application ExamplesSingle Channel Operation Commands DescriptionControl Page Sets the external interrupt output to be active high Bracketing a VoltageBracketing AN Input Voltage Page Register MAP Register Access ExamplesFor example Pseudo Register Access VXIPLUG&PLAY Driver Examples VM4016 Programming VXI Technology, Inc Alphabetical Command Listing Command DictionaryIeee 488.2 Common Commands Normal Instrument Specific Scpi CommandsScpi Required Commands Query Parameters Command DictionaryCLS Common Scpi CommandsResponse Description ESEESR? IDN? Query Parameters Query Response Description OPCRST SRE STB? TRG TST? OPC WAIFETChCONDitioned? Instrument Specific Scpi CommandsFETChLATChed? FETChRAW? Inhouseclearlatch Inhousepseudo Inhouseregint Inhouseregenable INPutDEBounce INPutMASK INPutMASKINTerrupt INPutOFFSet INPutPOLarity INPutRANGe OUTPutPOLarityEXTernalINTerrupt OUTPutPOLarityEXTernalLATChed STATusOPERationCONDition? Required Scpi CommandsSTATusOPERationENABle STATusOPERationEVENt? STATusPRESet STATusQUEStionableCONDition? STATusQUEStionableENABle STATusQUEStionableEVENt SYSTemERRor? SYSTemVERSion? Theory of Operation Input Range Control Relay Signal Comparison Interrupt Generation Interrupt Generation OUTPutPOLarityEXTernalLATChed Index