www.vxitech.com
|
| INPut:RANGe |
|
|
|
|
|
| Purpose | Sets the input range for one or more channels |
|
|
|
|
|
| Type | Setting |
|
|
|
|
|
| Command Syntax | INPut:RANGe <range>,<channel_list> |
|
|
|
|
|
| Command Parameters | <range> = 10 100 |
|
|
| <channel_list> = standard channel list syntax supporting channels 1 to 16 |
|
|
|
|
|
| *RST Value | 100 for all channels |
|
|
|
|
|
| Query Syntax | INPut:RANGe? <channel> |
|
|
|
|
|
| Query Parameters | <channel> = 1 to 16 |
|
|
|
|
|
| Query Response | Numeric ASCII value = 10 100 |
|
|
|
|
|
| Description | The Input Range command selects the input range of one or more channels. The input |
|
|
| range may be set for ±10 volts or ±100 volts. |
|
|
| Note: The input offset is normalized to the ±10 volt range. The actual input offset in |
|
|
| the 100 volt range is ten times the set value. |
|
|
|
|
|
|
|
|
|
| Examples | Command / Query | Response (Description) |
|
| INP:RANG 100,(@1,3,5,7) | (Sets the input range for Channels 1, 3, 5, and 7 to 100 |
|
|
| V) |
|
| INP:RANG? 7 | 100 (Returns the set input range for Channel 7) |
|
|
|
|
| Related Commands | INPut:OFFset <voltage_level>, | <channel_list> |
VM4016 Command Dictionary | 75 |