VXI Technology, Inc.
INPUT:POLARITY
The input polarity command selects the input polarity for one or more channels. When a channel is programmed for normal polarity, an interrupt will be generated when the input voltage is greater than the programmed input offset for the channel. The invert polarity will cause an interrupt when the input voltage is less than the programmed input offset for the channel.
INPut:POLarity <polarity>,<channel_list> | Where <polarity> is either NORMal or |
| INVerted |
| Where <channel_list> is the standard |
| channel list format supporting channels 1 |
| through 16. |
|
|
EXAMPLES |
|
INPut:POLarity NORM,(@3:5) | Sets the input polarity for Channels 3 |
| through 5 to NORMal. This will generate |
| an interrupt when the input signal on |
| Channels 3, 4 or 5 is greater than the input |
| offset. |
INPut:POLarity? 5 | Returns the input polarity for Channel 5 as |
NORM | NORMal |
INP:POL INV,(@6) | Sets input polarity for Channel 6 to invert. |
30 | VM4016 Programming |