Bunn S/N SNG0033000 & UP Bypass OZ/M, Calibrate Flow, Calibrate Spray .. Press Brew to Start

Page 34

PROGRAMMING THE BREWER (cont.)

SPRAY OZ/M (cont.)

SPRAY OZ/M: XXX (-) DONE (+)

P2721

Procedure to adjust the flow rate setting:

1.Press and hold the upper right hidden switch until the display reads UNITS. Release the switch, then press and release switch until the display reads SPRAY OZ/M. The number represents what the brewer thinks is the flow rate of the sprayhead valve in ounces per minute.

2.If the actual flow rate of the sprayhead valve is known, but is different than the number on the display, use the (-)and (+) to enter the correct flow rate in ounces per minute.

3.When finished, press and release DONE. This saves the new setting and advances to the next function screen, BYPASS OZ/M. Another alternative is to press the ON/OFF switch located on the switch panel to exit SPRAY OZ/M function and return to the MAIN SCREEN.

BYPASS OZ/M

This function allows the operator to view or to enter the actual flow rate coming out of the bypass nozzle. This is NOT used to change the actual flow rate, but to tell the internal controller how fast the water is flowing. The unit of measure is ounces per minute.

BYPAS OZ/M: XXX (-) DONE (+)

P2721

Procedure to adjust the flow rate setting:

1.Press and hold the upper right hidden switch until the display reads UNITS. Release the switch, then press and release switch until the display reads BYPAS OZ/M. The number represents what the brewer thinks is the flow rate of the bypass valve in ounces per minute.

2.If the actual flow rate of the bypass valve is known, but is different than the number on the display, use the (-)and (+) to enter the correct flow rate in ounces per minute.

3.When finished, press and release DONE. This saves the new setting and advances to the next function screen, CALIBRATE FLOW. Another alternative is to press the ON/OFF switch located on the front switch panel to exit the BYPASS OZ/M function and return to the MAIN SCREEN.

CALIBRATE FLOW

This function allows the operator to enter the actual flow rate of the sprayhead and the bypass by dispens- ing both separately for one minute. The volumes are then entered in ounces per minute. (Oz/M)

Procedures to calibrate the sprayhead flow rate:

1.Place a container, accurately graduated in ounces, and with a minimum capacity of 60 ounces, beneath the funnel on the brewer.

2.Press and hold the upper right hidden switch until the display reads UNITS. Release the switch, then press and release switch until the display reads

CALIBRATE FLOW.

3.Press and release YES to advance to the SPRAY- HEAD CAL? function screen. (Pressing NO in the CALIBRATE FLOW screen will advance to the next function screen, BREW COUNTERS).

4.Press and release the YES pad. The display should read CONTAINER READY? If the container is under the funnel, press YES. The display should read

CALIBRATE SPRAY .. PRESS BREW TO START.

5.Press and release the BREW switch. The display should read CALIBRATE SPRAY .. 60 SEC TO FIN- ISH. The 60 second timer on the display will count down to zero. When the counter reaches zero, the display will change to ENTER OZ., along with a number measured in ounces per minute.

6.Measure the amount of water in the container and using the (-)or (+), adjust the amount on the display to match the amount in the container. Then press DONE.

7.The display should now read NEW SPRAY FLOW, along with the correct flow rate of the sprayhead in ounces per minute. After about 5 seconds, the display will return to the CALIBRATE FLOW screen.

8.To exit the CALIBRATE FLOW program, press NO to advance to the next function screen, or press the ON/OFF switch located on the front switch panel to exit the CALIBRATE FLOW function and return to the MAIN SCREEN.

Page 34

35879.1 071405

 

Image 34
Contents Single TF DBC Phone 217 529-6601 FAX 217BUNN-O-MATIC Commercial Product Warranty Table of Contents User Notices Electrical HOOK-UP Volt ac single phase Hz models37881.0002 Electrical Requirements Models without power cord 230 volt ac Single phase, 60 Hz modelsPlumbing Requirements CE RequirementsPlumbing HOOK-UP ON/OFF Switch Enable Brew Main ON/OFF SwitchBatch Selector Switches Brew SwitchInitial Setup Coffee BrewingCleaning Glossary Programming Switches ProgrammingProgramming Switch Recipes Program Functions LevelDRIP-OUT Times Continues to D LevelReturns to a Return to a Program Flow DiagramContinues to Programming the Brewer Units Setting Units of Measure Procedure for modifying units of measureSET NEW Recipe Procedure for Setting the RecipeThis function allows the operator to set Brew Sizes DONE? SET NEW RecipesRefer to SET Pulse Brew for description Procedure to program the coffee name Card Contains Recipe forReview RECIPES? Noyes Next Review RECIPES/MODIFY RECIPES/SET UP no Name Coffee NamesProcedure for reviewing recipes To exit, press Next until the display reads ThatBrew OZ Setting or Adjusting Brew VOL- Umes ScreenBypass Review RecipesSET Pulse Brew Modifying pulse brew Range 1st on time off to 4 minutesNO-NAME Modifyshownext SET Pulse BrewSET Pulse Brew Easy Pulse Setup TIMES? SET Pulse Brew Enter TimesDONE? SET Preinfusion Batches DONE?Setup Complete Assign Recipes Enable ADS?Assign Recipe ? SET Ready SET TempScreen Spray OZ/M Calibrate Flow Bypass OZ/MCalibrate Spray .. Press Brew to Start Calibrate Flow Procedures to calibrate the bypass flow rateStart Brew Counters Procedure to set funnel detectProcedures to view/reset the brew counters Funnel Detect optionalProcedure to test components outputs Service ToolsTesting individual components outputs Brew ValveNothing Pressed FrequencyFactory Defaults Procedure to set factory defaultsScreen Displayed TroubleshootingTroubleshooting Procedures Possible CauseTemperature Press ON/OFF to terminate the cycleTank Heater failure ON/OFF switch was pressed after the brew cycle was startedPress ON/OFF to terminate the brew cycle Triac FailureWater shut off to brewer Temperature Sensor ProbeWires shorted to housing, or to Broken or not making connectionBrew Valve Switch test procedures Equipment will not operate No power or incorrect voltageBrew switch Switch test procedures Data plate within +/- 10%With power removed from brewer Holding it in open state Troubleshooting Problem Possible CauseOverflow Protection Switch When this condition occurs, See above or boilingLimit Thermostat Heating Time TOO LongTroubleshooting Problem Improper water pressure Inconsistent beverage level StrainerServer Pointing towards the brewer 90 psi 138 and 620 kPaLevel in server Consistently high or low beverage Brew Volume adjustmentWater temperature For each brew cycleBrewer is making unusual noises Troubleshooting Problem Possible Cause Weak beverageRemove and clean lime off tank heaters Plumbing lines Water Supply Tank HeatersSchematic Wiring Diagram Single TF DBC Schematic Wiring Diagram Wire Single Phase Schematic Wiring Diagram Single TF DBC