IMPORTANT SAFETY INSTRUCTIONS

In critical applications where very high requirements have to be met, it is strongly recommended that all the

relevant factors for the process, such as cleaning agent, quality of water etc. are discussed with the Miele Applications Specialists.

The mobile units and special inserts should only be used for

their specific application.

Empty any containers or utensils before arranging them in the

machine.

Do not allow any remains of acids or solvents, and in particular

hydrochloric acid, chloride solutions or corroding ferrous material to get into the wash cabinet. Similarly avoid any materials with a corrosive effect. The presence in compounds of any solvents should be minimal.

Ensure that solutions or steam containing hydrochloric acid do not

come into contact with the steel outer casing of the machine to avoid any corrosion damage.

Please follow the advice on installation in these instructions

and the separate Installation Instructions.

Using accessories

Only specific additional equipment made by Miele should be

connected to this machine. Consult the Miele Professional Group on the type and application of such equipment.

Disposal of an old appliance

When discarding a lab washer, disconnect it from the power

supply, remove the door to the wash cabinet and cut off the power cord. For environmental and safety reasons ensure the machine is completely drained of any residue water and cleaning agent. (Observe safety regulations and wear safety goggles and gloves). Make the door lock inoperative, so that children cannot accidentally shut themselves in. Make appropriate arrangements for the safe disposal of the machine.

If the machine contains a tank system, remove the water from the tank before disposing of the machine.

The manufacturer cannot be held responsible for any damage caused through non-observance of these Warning and Safety Instructions.

7

Image 7
Contents Laboratory glassware washer G Contents Contents Use Important Safety InstructionsImportant Safety Instructions Important Safety Instructions Disposal of an old appliance Using accessoriesElectronic controls Guide to the machineCursor right Door switch h Start buttonDisplay On/Off button Cursor leftGuide to the machine To close the door Opening and closing the doorElectric door lock Opening the doorWater softener Softener Capacity Setting the water softenerSelect Unit Enter Water HardnessFilling the salt reservoir Salt indicator Mobile unit coding Automatic mobile unit recognition AWKMobile unit coding strip Setting mobile unit codingSet the program place coding Removing excess soils Preparing the items for cleaningApplications Loading the machineApplications Example upper basket O 190/1 and lower basket U Height adjustable top basketTo adjust the upper basket Spring adapter water connectionInjector unit 1/1 With connection to TA Laboratory glassware LGPreparing the DOS dispensers Adding of liquid detergents and neutralizersAdditional DOS modules optional Filling the containers with liquid cleaning agents Optional Dispensing system maintenancePriming the liquid dispensing systems External Dispensing systemsTo change the operating level Power switchOperation Turning onFixed program Starting a programFree program selection Turning off Program sequenceAutomatic mobile unit recognition AWK Operating levels B or D To change or cancel a programOperating levels a or C Cancelling a program Serial interface Cleaning the filters in the wash cabinet Cleaning and CareCleaning the coarse filter Cleaning the fine, flat and micro-fine filters Tilt the locking clamp to the side Cleaning the drain pump and non-return valveCleaning the filter Cleaning the water inlet filtersCleaning the control panel Cleaning the exteriorChanging the coarse filter Drying unit TA Cleaning and careOpen the door of the storage cabinet Replacing the HEPA-Filter Frequently asked questions After sales service Page Installation Instructions Disposal of the packing materials Caring for the environmentDisposal of an old appliance Protecting the countertop depending on model InstallationLeveling the legs Removing the lid if necessaryWater inlet PlumbingDI-Water connection Drainage This Appliance Must be Grounded Electrical connectionGrounding the machine Re-assembling the service panelEnglish Metric Technical DataPage Page All rights reserved/3404