Teledyne Trace Oxygen Analyzer operating instructions Operation Model 3000TA-XL-EU

Page 40

4 Operation

Model 3000TA-XL-EU

 

 

 

 

 

 

 

 

 

After a password is assigned, the operator must log out to activate it. Until then, anyone can continue to operate the instrument without entering the new password.

Only one password can be defined. Before a unique password is assigned, the system assigns TETAI by default. This allows access to anyone. After a unique password is assigned, to defeat the security, the password must be changed back to TETAI.

Logout: Logging out prevents unauthorized tampering with analyzer settings.

More: Select and enter More to get a new screen with additional subfunctions listed.

Self–Test:The instrument performs a self-diagnostic test to check the integrity of the power supply, output boards and amplifiers.

Version: Displays Manufacturer, Model, and Software Version of instrument.

Show Negative: The operator selects whether display can show negative oxygen readings or not.

TRAK/HLD: The operator sets whether the instrument analog outputs track the concentration change during calibration and sets a time delay for the concentration alarms after calibration

4.3.1Tracking the Oxygen Readings during Calibration and Alarm delay

The user has the option of setting the preferenc as to whether the analog outputs track the display readings during calibration or not. To set the prefer- ence, press the System key once and the first System menu will appear in the VFD display:

TRAK/HLD Auto-Cal PSWD Logout More

TRAK/HLD should be blinking. To enter this system menu press the Enter key once:

Output Sttng: TRACK

Alarm Dly: 10 min Or

Output Sttng: HOLD

Alarm Dly: 10 min

4-4

Teledyne Analytical Instruments

Image 40
Contents Trace Oxygen Analyzer Teledyne Analytical InstrumentsModel 3000TA-XL-EU Copyright 1999 Teledyne Analytical InstrumentsIii Trace Oxygen Analyzer Specific Model InformationGround Protective Earth Teledyne Analytical InstrumentsTrace Oxygen Analyzer Table of Contents Introduction InstallationOperational Theory OperationMaintenance AppendixVii Trace Oxygen Analyzer AppendixCombustible GAS Usage Warning ViiiMain Features of the Analyzer OverviewTypical Applications Trace Oxygen Analyzer IntroductionModel Designations Introduction Model 3000TA-XL-EUFront Panel Operator Interface Model 3000TA-XL Front PanelIntroduction Model 3000TA-XL-EU Rear Panel Equipment Interface Analog OutputsRemote Probe RS-232 PortNetwork I/O Operational Theory Introduction Micro-Fuel Cell SensorTrace Oxygen Analyzer Operational Theory Principles of OperationOperational Theory Model 3000TA-XL-EU Anatomy of a Micro-Fuel CellElectrochemical Reactions + 2H 2O + 4e → 4OHCalibration Characteristics Effect of PressureCharacteristic Input/Output Curve for a Micro-Fuel Cell Sample SystemTOP Flow Diagram-Sample Under Pressure Standard Model Electronics and Signal Processing Electronic Component Location Inside the Model 3000TA-XLBlock Diagram of the Model 3000TA-XL Electronics Operational Theory Model 3000TA-XL-EU Unpacking the Analyzer Trace Oxygen Analyzer InstallationMounting the Analyzer Installation Model 3000TA-XL-EU Rear Panel ConnectionsGas Connections Rear Panel of the Model 3000TA-XL2.2 50-Pin Equipment Interface Connector PrimaryInputPowerElectrical Connections Equipment Interface Connector Pin Arrangement Pin FunctionPin Contact Remote Calibration Connections Range ID Relay Connections 2.3 RS-232 Port FET Series ResistanceCommand Description Parameter SettingInstalling the Micro-Fuel Cell Testing the SystemRemoving the Micro-Fuel Cell Trace Oxygen Analyzer Operation Auto Ranging onUsing the Data Entry and Function Buttons Operation Model 3000TA-XL-EUSystem Function Analyze SystemOperation Model 3000TA-XL-EU Setting up an Auto-Cal Password Protection Entering the Password Installing or Changing the Password System Self-Diagnostic Test LogoutVersion Screen Calibration of the Analyzer Span Gas CalibrationAir Calibration Zero CalAuto Mode Zeroing Manual Mode ZeroingCell Failure Auto Mode Spanning Operation Model 3000TA-XL-EU Span CalManual Mode Spanning Switching of Sample Streams Span FailureAlarms Function Special Notes on Hydrogen Gas StreamAL-1 AL-2 Choose Alarm Range Function Setting the Analog Output Ranges Fixed Range AnalysisSignal Output Analyze FunctionRange Voltage Current mA MEDRoutine Maintenance Cell ReplacementTrace Oxygen Analyzer Maintenance Storing and Handling Replacement CellsMaintenance Model 3000TA-XL-EU When to Replace a Cell Removing the Micro-Fuel CellRemoving the Micro-Fuel Cell Warranty Fuse Replacement Removing Fuse Block from HousingMaintenance Model 3000TA-XL-EU System Self Diagnostic TestMajor Internal Components Cleaning Rear-Panel ScrewsTroubleshooting SolutionSpecifications Operating Temperature 5-35 C AppendixRecommended 2-Year Spare Parts List Drawing List 19-inch Relay Rack Panel Mount3000XL Series Analyzers Conversions Page Material Safety Data Sheet Section IV Fire and Explosion Hazard Data Protective measures during cell replacement ExposureSection Viii Control Measures Read completely before beginning the procedure