Teledyne 3000TA-XL-EU, Trace Oxygen Analyzer operating instructions Specifications

Page 67

Trace Oxygen Analyzer

Appendix

 

 

 

 

 

 

 

 

 

Appendix

A-1 Specifications

Packaging:

General Purpose

 

• Flush panel mount (Standard).

 

• Relay rack mount. Contains either one or

 

two instruments in one 19" relay rack

 

mountable plate (Optional).

Sensor:

Teledyne B-2CXL trace analysis Micro-Fuel

 

Cell.

Cell Block:

316 stainless steel.

Sample System:

All wetted parts of 316 stainless steel.

90 % Response Time:

65 seconds at 25 °C (77 °F) on 10 ppm range

 

100 seconds on 1 ppm range

Ranges:

Three user definable ranges from 0–1 ppm to 0–

 

250,000 ppm, plus air calibration range of 0-

 

250,000 ppm (25 %).

 

Autoranging with range ID output.

Alarms:

One system-failure alarm contact to detect

 

power failure or sensor-zero failure.

 

Two adjustable concentration threshold alarm

 

contacts with fully programmable setpoints.

Displays:

2-line by 20-character, VFD screen, and one 5

 

digit LED display.

Digital Interface:

Full duplex RS-232 communications port.

Teledyne Analytical Instruments

A-1

Image 67
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 1999 Teledyne Analytical Instruments Model 3000TA-XL-EUTrace Oxygen Analyzer Specific Model Information IiiProtective Earth Teledyne Analytical Instruments GroundOperation Trace Oxygen Analyzer Table of Contents IntroductionInstallation Operational TheoryAppendix MaintenanceTrace Oxygen Analyzer Appendix ViiViii Combustible GAS Usage WarningTrace Oxygen Analyzer Introduction Main Features of the AnalyzerOverview Typical ApplicationsIntroduction Model 3000TA-XL-EU Model DesignationsModel 3000TA-XL Front Panel Front Panel Operator InterfaceIntroduction Model 3000TA-XL-EU Analog Outputs Rear Panel Equipment InterfaceRemote Probe RS-232 PortNetwork I/O Principles of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Trace Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model 3000TA-XL-EU+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsSample System Characteristic Input/Output Curve for a Micro-Fuel CellTOP Flow Diagram-Sample Under Pressure Standard Model Electronic Component Location Inside the Model 3000TA-XL Electronics and Signal ProcessingBlock Diagram of the Model 3000TA-XL Electronics Operational Theory Model 3000TA-XL-EU Unpacking the Analyzer Trace Oxygen Analyzer InstallationMounting the Analyzer Rear Panel Connections Installation Model 3000TA-XL-EURear Panel of the Model 3000TA-XL Gas Connections2.2 50-Pin Equipment Interface Connector PrimaryInputPowerElectrical Connections Pin Function Equipment Interface Connector Pin ArrangementPin Contact Remote Calibration Connections Range ID Relay Connections FET Series Resistance 2.3 RS-232 PortParameter Setting Command DescriptionTesting the System Installing the Micro-Fuel CellRemoving the Micro-Fuel Cell Auto Ranging on Trace Oxygen Analyzer OperationOperation Model 3000TA-XL-EU Using the Data Entry and Function ButtonsAnalyze System System FunctionOperation Model 3000TA-XL-EU Setting up an Auto-Cal Password Protection Entering the Password Installing or Changing the Password Logout System Self-Diagnostic TestVersion Screen Zero Cal Calibration of the AnalyzerSpan Gas Calibration Air CalibrationManual Mode Zeroing Auto Mode ZeroingCell Failure Operation Model 3000TA-XL-EU Span Cal Auto Mode SpanningManual Mode Spanning Span Failure Switching of Sample StreamsSpecial Notes on Hydrogen Gas Stream Alarms FunctionAL-1 AL-2 Choose Alarm Range Function Fixed Range Analysis Setting the Analog Output RangesAnalyze Function Signal OutputMED Range Voltage Current mAStoring and Handling Replacement Cells Routine MaintenanceCell Replacement Trace Oxygen Analyzer MaintenanceRemoving the Micro-Fuel Cell Maintenance Model 3000TA-XL-EU When to Replace a CellRemoving the Micro-Fuel Cell Warranty Removing Fuse Block from Housing Fuse ReplacementMaintenance Model 3000TA-XL-EU System Self Diagnostic TestMajor Internal Components Rear-Panel Screws CleaningSolution TroubleshootingSpecifications Appendix Operating Temperature 5-35 CRecommended 2-Year Spare Parts List 19-inch Relay Rack Panel Mount Drawing List3000XL Series Analyzers Conversions Page Material Safety Data Sheet Section IV Fire and Explosion Hazard Data Exposure Protective measures during cell replacementSection Viii Control Measures Read completely before beginning the procedure
Related manuals
Manual 17 pages 27.99 Kb