Teledyne 3000TA-XL-EU, Trace Oxygen Analyzer operating instructions Manual Mode Spanning

Page 51

Trace Oxygen Analyzer

Operation 4

 

 

 

 

 

 

 

 

 

sponds, within tolerance, to the value of the span gas concentration. Then the instrument automatically returns to the analyze mode.

4.4.2.2Manual Mode Spanning

Press Span to start the Span function. The screen that appears allows you to select whether the span calibration is to be performed automatically or manually.

Span:Settling:MAN

<ENT> For Next

Use the Δ∇ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen.

Press Enter to move to the next screen.

Calib.Holdingtime

Cal hold: 5 min

This menue allows the operator to set the time the analyzer should be held in the AUTO span mode. It does not have any effect in the MANual mode. Just press Enter key to continue.

Span Val: 000008.00 <ENT>Span <UP>Mod #

Press Δ (<UP>) to permit modification (Mod #) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas you are using (209000.00 if you are using air). The < > arrows choose the digit, and the Δ∇ arrows choose the value of the digit.

Press Enter to enter the span value into the system and begin the span calibration.

Once the span has begun, the microprocessor samples the output at a predetermined rate. It calculates the difference between successive samplings and displays this difference as a Slope on the screen. It takes several seconds for the first Slope value to display. Slope indicates the rate of change of the Span reading. It is a sensitive indicator of stability.

####% Span

Slope=#### ppm/s

When the Span value displayed on the screen is sufficiently stable, press Enter. (Generally, when the Span reading changes by 1 % or less of the full scale of the range being calibrated, for a period of ten minutes it is

Teledyne Analytical Instruments

4-15

Image 51
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 1999 Teledyne Analytical Instruments Model 3000TA-XL-EUTrace Oxygen Analyzer Specific Model Information IiiProtective Earth Teledyne Analytical Instruments GroundOperation Trace Oxygen Analyzer Table of Contents IntroductionInstallation Operational TheoryAppendix MaintenanceTrace Oxygen Analyzer Appendix ViiViii Combustible GAS Usage WarningTrace Oxygen Analyzer Introduction Main Features of the AnalyzerOverview Typical ApplicationsIntroduction Model 3000TA-XL-EU Model DesignationsModel 3000TA-XL Front Panel Front Panel Operator InterfaceIntroduction Model 3000TA-XL-EU Analog Outputs Rear Panel Equipment InterfaceRS-232 Port Remote ProbeNetwork I/O Principles of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Trace Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model 3000TA-XL-EU+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsSample System Characteristic Input/Output Curve for a Micro-Fuel CellTOP Flow Diagram-Sample Under Pressure Standard Model Electronic Component Location Inside the Model 3000TA-XL Electronics and Signal ProcessingBlock Diagram of the Model 3000TA-XL Electronics Operational Theory Model 3000TA-XL-EU Trace Oxygen Analyzer Installation Unpacking the AnalyzerMounting the Analyzer Rear Panel Connections Installation Model 3000TA-XL-EURear Panel of the Model 3000TA-XL Gas ConnectionsPrimaryInputPower 2.2 50-Pin Equipment Interface ConnectorElectrical Connections Pin Function Equipment Interface Connector Pin ArrangementPin Contact Remote Calibration Connections Range ID Relay Connections FET Series Resistance 2.3 RS-232 PortParameter Setting Command DescriptionTesting the System Installing the Micro-Fuel CellRemoving the Micro-Fuel Cell Auto Ranging on Trace Oxygen Analyzer OperationOperation Model 3000TA-XL-EU Using the Data Entry and Function ButtonsAnalyze System System FunctionOperation Model 3000TA-XL-EU Setting up an Auto-Cal Password Protection Entering the Password Installing or Changing the Password Logout System Self-Diagnostic TestVersion Screen Zero Cal Calibration of the AnalyzerSpan Gas Calibration Air CalibrationManual Mode Zeroing Auto Mode ZeroingCell Failure Operation Model 3000TA-XL-EU Span Cal Auto Mode SpanningManual Mode Spanning Span Failure Switching of Sample StreamsSpecial Notes on Hydrogen Gas Stream Alarms FunctionAL-1 AL-2 Choose Alarm Range Function Fixed Range Analysis Setting the Analog Output RangesAnalyze Function Signal OutputMED Range Voltage Current mAStoring and Handling Replacement Cells Routine MaintenanceCell Replacement Trace Oxygen Analyzer MaintenanceRemoving the Micro-Fuel Cell Maintenance Model 3000TA-XL-EU When to Replace a CellRemoving the Micro-Fuel Cell Warranty Removing Fuse Block from Housing Fuse ReplacementSystem Self Diagnostic Test Maintenance Model 3000TA-XL-EUMajor Internal Components Rear-Panel Screws CleaningSolution TroubleshootingSpecifications Appendix Operating Temperature 5-35 CRecommended 2-Year Spare Parts List 19-inch Relay Rack Panel Mount Drawing List3000XL Series Analyzers Conversions Page Material Safety Data Sheet Section IV Fire and Explosion Hazard Data Exposure Protective measures during cell replacementSection Viii Control Measures Read completely before beginning the procedure