Teledyne Trace Oxygen Analyzer, 3000TA-XL-EU operating instructions AL-1 AL-2 Choose Alarm

Page 54

4 Operation

Model 3000TA-XL-EU

 

 

 

 

 

 

 

 

 

Both high (high and high-high) alarms, or

One high and one low alarm, or

Both low (low and low-low) alarms.

2.Are either or both of the alarms to be configured as failsafe?

In failsafe mode, the alarm relay de-energizes in an alarm condition. For non-failsafe operation, the relay is energized in an alarm condition. You can set either or both of the concentration alarms to operate in failsafe or non-failsafe mode.

3.Are either of the alarms to be latching?

In latching mode, once the alarm or alarms trigger, they will remain in the alarm mode even if process conditions revert back to non-alarm conditions. This mode requires an alarm to be recognized before it can be reset. In the non-latching mode, the alarm status will terminate when process conditions revert to non- alarm conditions.

4.Are either of the alarms to be defeated?

The defeat alarm mode is incorporated into the alarm circuit so that maintenance can be performed under conditions which would normally activate the alarms.

The defeat function can also be used to reset a latched alarm. (See procedures, below.)

If you are using password protection, you will need to enter your password to access the alarm functions. Follow the instructions in section

4.3.3to enter your password. Once you have clearance to proceed, enter the Alarm function.

Press the Alarm button on the front panel to enter the Alarm function. Make sure that AL–1is blinking.

AL—1 AL—2

Choose Alarm

Set up alarm 1 by moving the blinking over to AL–1 using the < > arrow keys. Then press Enter to move to the next screen.

AL—1 1000 ppm HI

Dft—N Fs—N Ltch—N

Five parameters can be changed on this screen:

• Value of the alarm setpoint, AL–1 #### ppm (oxygen); value can be set from 0.00 to 10,000.00 ppm

Out-of-range direction, HI or LO

4-18

Teledyne Analytical Instruments

Image 54
Contents Trace Oxygen Analyzer Teledyne Analytical InstrumentsModel 3000TA-XL-EU Copyright 1999 Teledyne Analytical InstrumentsIii Trace Oxygen Analyzer Specific Model InformationGround Protective Earth Teledyne Analytical InstrumentsOperational Theory Trace Oxygen Analyzer Table of Contents IntroductionInstallation OperationMaintenance AppendixVii Trace Oxygen Analyzer AppendixCombustible GAS Usage Warning ViiiTypical Applications Main Features of the AnalyzerOverview Trace Oxygen Analyzer IntroductionModel Designations Introduction Model 3000TA-XL-EUFront Panel Operator Interface Model 3000TA-XL Front PanelIntroduction Model 3000TA-XL-EU Rear Panel Equipment Interface Analog OutputsRS-232 Port Remote ProbeNetwork I/O Trace Oxygen Analyzer Operational Theory Operational Theory IntroductionMicro-Fuel Cell Sensor Principles of OperationOperational Theory Model 3000TA-XL-EU Anatomy of a Micro-Fuel CellElectrochemical Reactions + 2H 2O + 4e → 4OHCalibration Characteristics Effect of PressureCharacteristic Input/Output Curve for a Micro-Fuel Cell Sample SystemTOP Flow Diagram-Sample Under Pressure Standard Model Electronics and Signal Processing Electronic Component Location Inside the Model 3000TA-XLBlock Diagram of the Model 3000TA-XL Electronics Operational Theory Model 3000TA-XL-EU Trace Oxygen Analyzer Installation Unpacking the AnalyzerMounting the Analyzer Installation Model 3000TA-XL-EU Rear Panel ConnectionsGas Connections Rear Panel of the Model 3000TA-XLPrimaryInputPower 2.2 50-Pin Equipment Interface ConnectorElectrical Connections Equipment Interface Connector Pin Arrangement Pin FunctionPin Contact Remote Calibration Connections Range ID Relay Connections 2.3 RS-232 Port FET Series ResistanceCommand Description Parameter SettingInstalling the Micro-Fuel Cell Testing the SystemRemoving the Micro-Fuel Cell Trace Oxygen Analyzer Operation Auto Ranging onUsing the Data Entry and Function Buttons Operation Model 3000TA-XL-EUSystem Function Analyze SystemOperation Model 3000TA-XL-EU Setting up an Auto-Cal Password Protection Entering the Password Installing or Changing the Password System Self-Diagnostic Test LogoutVersion Screen Air Calibration Calibration of the AnalyzerSpan Gas Calibration Zero CalAuto Mode Zeroing Manual Mode ZeroingCell Failure Auto Mode Spanning Operation Model 3000TA-XL-EU Span CalManual Mode Spanning Switching of Sample Streams Span FailureAlarms Function Special Notes on Hydrogen Gas StreamAL-1 AL-2 Choose Alarm Range Function Setting the Analog Output Ranges Fixed Range AnalysisSignal Output Analyze FunctionRange Voltage Current mA MEDTrace Oxygen Analyzer Maintenance Routine MaintenanceCell Replacement Storing and Handling Replacement CellsMaintenance Model 3000TA-XL-EU When to Replace a Cell Removing the Micro-Fuel CellRemoving the Micro-Fuel Cell Warranty Fuse Replacement Removing Fuse Block from HousingSystem Self Diagnostic Test Maintenance Model 3000TA-XL-EUMajor Internal Components Cleaning Rear-Panel ScrewsTroubleshooting SolutionSpecifications Operating Temperature 5-35 C AppendixRecommended 2-Year Spare Parts List Drawing List 19-inch Relay Rack Panel Mount3000XL Series Analyzers Conversions Page Material Safety Data Sheet Section IV Fire and Explosion Hazard Data Protective measures during cell replacement ExposureSection Viii Control Measures Read completely before beginning the procedure