Trace Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
Wait a moment for the entry screen. You will be given clearance to proceed.
A A A A A <ENT> TO Proceed
Use the arrow keys to retype your password and press Enter when finished. Your password will be stored in the microprocessor and the system will immediately switch to the Analyze screen, and you now have access to all instrument functions.
If all alarms are defeated, the Analyze screen appears as:
0.0ppm Anlz Range: 0 — 100
If an alarm is tripped, the second line will change to show which alarm
it is:
0.0ppm Anlz
AL — 1
NOTE: If you log off the system using the logout function in the system menu, you will now be required to
4.3.4 Logout
The Logout function provides a convenient means of leaving the analyzer in a password protected mode without having to shut the instrument off. By entering Logout, you effectively log off the instrument leaving the system protected against use until the password is reentered. To log out, press the System button to enter the System function.
TRAK/HLD
PSWD Logout More
Use the < > arrow keys to position the blinking over the Logout func- tion, and press Enter to Log out. The screen will display the message:
ProtectedUntil
Password Reentered
4.3.5 System Self-Diagnostic Test
The Model
Teledyne Analytical Instruments |