4 Operation | Model | |
|
|
|
|
|
|
|
|
|
status of each function is displayed on the screen, either as OK or as a number between 1 and 3. (See System Self Diagnostic Test in chapter 5 for number code.)
The self diagnostics are run automatically by the analyzer whenever the instrument is turned on, but the test can also be run by the operator at will. To initiate a self diagnostic test during operation:
Press the System button to start the System function.
TRAK/HLD
PSWD Logout More
Use the < > arrow keys to blink More, then press Enter.
Use the < > arrow keys again to move the blinking to the
RUNNING DIAGNOSTIC
Testing Preamp — 83
During preamp testing there is a countdown in the lower right corner of the screen. When the testing is complete, the results are displayed.
Power: OK Analog: OK
Preamp: 3
The module is functioning properly if it is followed by OK. A number indicates a problem in a specific area of the instrument. Refer to Chapter 5 Maintenance and Troubleshooting for
Press Any Key
ToContinue...
Then the analyzer returns to the initial System screen.
4.3.6 Version Screen
Move the < > arrow key to More and press Enter. With Version blinking, press Enter. The screen displays the manufacturer, model, and software version information.
Teledyne Analytical Instruments |