Trace Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
nitrogen and helium will not require SPAN adjustment. However, care must be taken to keep sample flow rates in the desired levels, as the flowmeter is calibrated for air and nitrogen streams only.
3.The analyzer can withstand repeated switching of streams between low oxygen levels of less than 1 ppm and high oxygen levels of 5000 to 10,000 ppm. Again, care must be taken to ensure proper and continuous flows at all times.
4.5.1Special Notes on Hydrogen Gas Stream
The standard oxygen cell, Model
4.6The Alarms Function
The Model
The system failure alarm has a fixed configuration as described in chapter 3 Installation.
The concentration alarms can be configured from the front panel as either high or low alarms by the operator. The alarm modes can be set as latching or
Decide how your alarms should be configured. The choice will depend upon your process. Consider the following four points:
1.Which if any of the alarms are to be high alarms and which if any are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the oxygen concentration rises above the setpoint. Setting an alarm as LOW triggers the alarm when the oxygen concentration falls below the setpoint.
Decide whether you want the alarms to be set as:
Teledyne Analytical Instruments |