Trace Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
•Defeated?
•Failsafe?
•Latching?
•To define the setpoint, use the < > arrow keys to move the blinking over to
•To set the other parameters use the < > arrow keys to move the blinking over to the desired parameter. Then use the Δ∇ arrow keys to change the parameter.
•Once the parameters for alarm 1 have been set, press Alarms again, and repeat this procedure for alarm 2
•To reset a latched alarm, go to Dft– and then press either Δ two times or ∇ two times. (Toggle it to Y and then back to N.)
Go to Ltch– and then press either Δ two times or ∇ two times. (Toggle it to N and back to Y.)
4.7The Range Function
The Range function allows the operator to program up to three concen- tration ranges to correlate with the DC analog outputs. If no ranges are defined by the user, the instrument defaults to:
Low =
The Model
The autoranging feature can be overridden so that analog output stays on a fixed range regardless of the oxygen concentration detected. If the
Teledyne Analytical Instruments |