Teledyne Trace Oxygen Analyzer, 3000TA-XL-EU Command Description, Parameter Setting

Page 34

3 Installation

Model 3000TA-XL-EU

 

 

 

 

 

 

 

 

 

 

Table 3-5: Commands via RS-232 Input

Command

Description

as<enter>

Immediately starts an autospan.

az<enter>

Immediately starts an autozero.

st<enter>

Toggling input. Stops/Starts any status message output from

 

the RS-232, until st<enter> is sent again.

The RS-232 protocol allows some flexibility in its implementation. Table 3-6 lists certain RS-232 values that are required by the 3000TA implementation.

Table 3-6: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 secondsWhen CRC opens again,

zero and span are done, and the sample is being analyzed.

Range ID Relays: There are four dedicated Range ID relay contacts. The first three ranges are assigned to relays in ascending order—Low range is assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and High range is assigned to Range 3 ID. The fourth range is reserved for the Air Cal Range (25%).

Network I/O: A serial digital input/output for local network protocol. At this printing, this port is not yet functional. It is to be used in future options to the instrument.

RS-232 Port: The digital signal output is a standard RS-232 serial communications port used to connect the analyzer to a computer, terminal, or other digital device. It requires a standard 9-pin D connector.

The data is status information, in digital form, updated every two seconds. Status is reported in the following order:

The concentration in ppm or percent

The range in use (HI, MED, LO)

The span of the range (0-100 ppm, etc)

Which alarms—if any—are disabled (AL–x DISABLED)

Which alarms—if any—are tripped (AL–x ON).

3-10

Teledyne Analytical Instruments

Image 34
Contents Trace Oxygen Analyzer Teledyne Analytical InstrumentsModel 3000TA-XL-EU Copyright 1999 Teledyne Analytical InstrumentsIii Trace Oxygen Analyzer Specific Model InformationGround Protective Earth Teledyne Analytical InstrumentsOperational Theory Trace Oxygen Analyzer Table of Contents IntroductionInstallation OperationMaintenance AppendixVii Trace Oxygen Analyzer AppendixCombustible GAS Usage Warning ViiiTypical Applications Main Features of the AnalyzerOverview Trace Oxygen Analyzer IntroductionModel Designations Introduction Model 3000TA-XL-EUFront Panel Operator Interface Model 3000TA-XL Front PanelIntroduction Model 3000TA-XL-EU Rear Panel Equipment Interface Analog OutputsRemote Probe RS-232 PortNetwork I/O Trace Oxygen Analyzer Operational Theory Operational Theory IntroductionMicro-Fuel Cell Sensor Principles of OperationOperational Theory Model 3000TA-XL-EU Anatomy of a Micro-Fuel CellElectrochemical Reactions + 2H 2O + 4e → 4OHCalibration Characteristics Effect of PressureCharacteristic Input/Output Curve for a Micro-Fuel Cell Sample SystemTOP Flow Diagram-Sample Under Pressure Standard Model Electronics and Signal Processing Electronic Component Location Inside the Model 3000TA-XLBlock Diagram of the Model 3000TA-XL Electronics Operational Theory Model 3000TA-XL-EU Unpacking the Analyzer Trace Oxygen Analyzer InstallationMounting the Analyzer Installation Model 3000TA-XL-EU Rear Panel ConnectionsGas Connections Rear Panel of the Model 3000TA-XL2.2 50-Pin Equipment Interface Connector PrimaryInputPowerElectrical Connections Equipment Interface Connector Pin Arrangement Pin FunctionPin Contact Remote Calibration Connections Range ID Relay Connections 2.3 RS-232 Port FET Series ResistanceCommand Description Parameter SettingInstalling the Micro-Fuel Cell Testing the SystemRemoving the Micro-Fuel Cell Trace Oxygen Analyzer Operation Auto Ranging onUsing the Data Entry and Function Buttons Operation Model 3000TA-XL-EUSystem Function Analyze SystemOperation Model 3000TA-XL-EU Setting up an Auto-Cal Password Protection Entering the Password Installing or Changing the Password System Self-Diagnostic Test LogoutVersion Screen Air Calibration Calibration of the AnalyzerSpan Gas Calibration Zero CalAuto Mode Zeroing Manual Mode ZeroingCell Failure Auto Mode Spanning Operation Model 3000TA-XL-EU Span CalManual Mode Spanning Switching of Sample Streams Span FailureAlarms Function Special Notes on Hydrogen Gas StreamAL-1 AL-2 Choose Alarm Range Function Setting the Analog Output Ranges Fixed Range AnalysisSignal Output Analyze FunctionRange Voltage Current mA MEDTrace Oxygen Analyzer Maintenance Routine MaintenanceCell Replacement Storing and Handling Replacement CellsMaintenance Model 3000TA-XL-EU When to Replace a Cell Removing the Micro-Fuel CellRemoving the Micro-Fuel Cell Warranty Fuse Replacement Removing Fuse Block from HousingMaintenance Model 3000TA-XL-EU System Self Diagnostic TestMajor Internal Components Cleaning Rear-Panel ScrewsTroubleshooting SolutionSpecifications Operating Temperature 5-35 C AppendixRecommended 2-Year Spare Parts List Drawing List 19-inch Relay Rack Panel Mount3000XL Series Analyzers Conversions Page Material Safety Data Sheet Section IV Fire and Explosion Hazard Data Protective measures during cell replacement ExposureSection Viii Control Measures Read completely before beginning the procedure