Teledyne 3000TA-XL-EU, Trace Oxygen Analyzer 2.3 RS-232 Port, FET Series Resistance

Page 33

Trace Oxygen Analyzer

Installation 3

 

 

 

 

 

 

 

 

 

The voltage from these outputs is nominally 0 V for the OFF and

15 V dc for the ON conditions. The maximum combined current that can be pulled from these output lines is 100 mA. (If two lines are ON at the same time, each must be limited to 50 mA, etc.) If more current and/or a different voltage is required, use a relay, power amplifier, or other matching circuitry to provide the actual driving current.

In addition, each individual line has a series FET with a nominal ON resistance of 5 ohms (9 ohms worst case). This can limit the obtainable

voltage, depending on the load impedance applied. See Figure 3-6

.

Figure 3-6: FET Series Resistance

3.3.2.3RS-232 Port

The digital signal output is a standard, full duplex RS-232 serial com- munications port used to connect the analyzer to a computer, terminal, or other digital device. It requires a standard 9-pin D connector.

The output data is status information, in digital form, updated every two seconds. Status is reported in the following order:

The concentration in ppm or percent

The range in use (HI, MED, LO)

The span of the range (0-100 ppm, etc)

Which alarms—if any—are disabled (AL–x DISABLED)

Which alarms—if any—are tripped (AL–x ON).

Each status output is followed by a carriage return and line feed.

Three input functions using RS-232 have been implemented to date. They are described in Table 3-5.

Teledyne Analytical Instruments

3-9

Image 33
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 1999 Teledyne Analytical Instruments Model 3000TA-XL-EUTrace Oxygen Analyzer Specific Model Information IiiProtective Earth Teledyne Analytical Instruments GroundInstallation Trace Oxygen Analyzer Table of Contents IntroductionOperational Theory OperationAppendix MaintenanceTrace Oxygen Analyzer Appendix ViiViii Combustible GAS Usage WarningOverview Main Features of the AnalyzerTypical Applications Trace Oxygen Analyzer IntroductionIntroduction Model 3000TA-XL-EU Model DesignationsModel 3000TA-XL Front Panel Front Panel Operator InterfaceIntroduction Model 3000TA-XL-EU Analog Outputs Rear Panel Equipment InterfaceRS-232 Port Remote ProbeNetwork I/O Micro-Fuel Cell Sensor Operational Theory IntroductionTrace Oxygen Analyzer Operational Theory Principles of OperationAnatomy of a Micro-Fuel Cell Operational Theory Model 3000TA-XL-EU+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsSample System Characteristic Input/Output Curve for a Micro-Fuel CellTOP Flow Diagram-Sample Under Pressure Standard Model Electronic Component Location Inside the Model 3000TA-XL Electronics and Signal ProcessingBlock Diagram of the Model 3000TA-XL Electronics Operational Theory Model 3000TA-XL-EU Trace Oxygen Analyzer Installation Unpacking the AnalyzerMounting the Analyzer Rear Panel Connections Installation Model 3000TA-XL-EURear Panel of the Model 3000TA-XL Gas ConnectionsPrimaryInputPower 2.2 50-Pin Equipment Interface ConnectorElectrical Connections Pin Function Equipment Interface Connector Pin ArrangementPin Contact Remote Calibration Connections Range ID Relay Connections FET Series Resistance 2.3 RS-232 PortParameter Setting Command DescriptionTesting the System Installing the Micro-Fuel CellRemoving the Micro-Fuel Cell Auto Ranging on Trace Oxygen Analyzer OperationOperation Model 3000TA-XL-EU Using the Data Entry and Function ButtonsAnalyze System System FunctionOperation Model 3000TA-XL-EU Setting up an Auto-Cal Password Protection Entering the Password Installing or Changing the Password Logout System Self-Diagnostic TestVersion Screen Span Gas Calibration Calibration of the AnalyzerAir Calibration Zero CalManual Mode Zeroing Auto Mode ZeroingCell Failure Operation Model 3000TA-XL-EU Span Cal Auto Mode SpanningManual Mode Spanning Span Failure Switching of Sample StreamsSpecial Notes on Hydrogen Gas Stream Alarms FunctionAL-1 AL-2 Choose Alarm Range Function Fixed Range Analysis Setting the Analog Output RangesAnalyze Function Signal OutputMED Range Voltage Current mACell Replacement Routine MaintenanceTrace Oxygen Analyzer Maintenance Storing and Handling Replacement CellsRemoving the Micro-Fuel Cell Maintenance Model 3000TA-XL-EU When to Replace a CellRemoving the Micro-Fuel Cell Warranty Removing Fuse Block from Housing Fuse ReplacementSystem Self Diagnostic Test Maintenance Model 3000TA-XL-EUMajor Internal Components Rear-Panel Screws CleaningSolution TroubleshootingSpecifications Appendix Operating Temperature 5-35 CRecommended 2-Year Spare Parts List 19-inch Relay Rack Panel Mount Drawing List3000XL Series Analyzers Conversions Page Material Safety Data Sheet Section IV Fire and Explosion Hazard Data Exposure Protective measures during cell replacementSection Viii Control Measures Read completely before beginning the procedure