Teledyne Percent Oxygen Analyzer, 9060L operating instructions Zirconia Sensor

Page 18

Operational Theory

Model 9060L

 

 

 

2.2.2 The Zirconia Sensor

Input to the analyzer is obtained from a solid electrolyte oxygen probe containing a zirconia element and thermocouple. The probe is designed to be inserted into a boiler or furnace exit gas flue or similar process. The 9060LEX sensor is designed to be installed outside of the flue or process. Sampling lines and filters are not required for in-situ probes but they are required for 9060LEX sensors. When a sampling line is required, the gas typically flows to the sensor under process pressure. In applications where the process pressure is negative or neutral, a suction pump will be required. A reference air pump is not provided in the Model 9060L Oxygen Analyzer. The internal construction of a probe or sensor is shown in Figure 2-3.

Figure 2-3: Internal Structure of the Zirconia Sensor and Probe

Heater control in the Model 9060L Oxygen Analyzer consists of a time proportioning temperature controller and solid state relay so that the thermocouple junction is controller to over 700°C (1300°F). Probes operating in a combustion environment above 650°C (1200°F) do not

Teledyne Analytical Instruments

18

Image 18
Contents Model 9060L Teledyne Analytical InstrumentsModel 9060L Copyright 2005 Teledyne Instruments/ Analytical InstrumentsSpecific Model Information Percent Oxygen AnalyzerSafety Messages Percent Oxygen Analyzer Table of Contents Appendix MaintenanceOperation Index Teledyne Analytical InstrumentsList of Figures Front PanelList of Tables Combustible GAS Usage Overview Main Features of the AnalyzerIntroduction Percent Oxygen Analyzer IntroductionFront Panel Description Introduction Model 9060LRear Panel Description SpanRear Panel Oxygen Probe, Sensor, and Heater Operational TheoryIntroduction Percent Oxygen Analyzer Operational TheoryOperational Theory Model 9060L Model 9060L Oxygen ProbeProbe Configuration Probe OptionsZirconia Sensor Internal Structure of the Zirconia Sensor and ProbeHeater Electronics Applications Not at Atmospheric PressureSensor Impedance GeneralSignal Processing Signal Processing Block DiagramOperational Theory Model 9060L 9060L Oxygen Probe Installation InstallationControl Unit Installation Percent Oxygen Analyzer InstallationInstallation Model 9060L Primary Input Power Electrical ConnectionsAlarm Relays Analog OutputsAlarm #2 Oxygen Probe Connections 4Model 9060L Heated Probe Cable ConnectionInstallation Checklist Connections to ProbeInstallation Model 9060L Teledyne Analytical Instruments Operation Percent Oxygen Analyzer OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation Model 9060L HI Range Setting the Alarm SetpointsSet Alarm 1 HI alarm LO RangeSet Alarm 2 LO alarm Sensor Fail AlarmSelecting a Fixed Range or Autoranging Oxygen Calibration Span CalibrationOperation Model 9060L Probe Temperature Supplementary InformationAir Calibration Error Messages Low TemperatureOpen Thermocouple Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Maintenance Percent Oxygen Analyzer MaintenanceReplacing the Fuse Maintenance Model 9060L Specifications AppendixProbe Specifications Appendix Model 9060LSpare Parts List Email asktai@teledyne.comAnalytical Instruments Reference Drawing Probe or Sensor EMF TablesZirconia Oxygen Sensor Output mV Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer