Percent Oxygen Analyzer | Maintenance | |
|
|
|
Maintenance
Aside from normal cleaning and checking for leaks at the probe connection, the Model 9060L should not require any maintenance beyond replacement of a blown fuse. Routine maintenance includes occasional recalibration, as described in Chapter 4, Operation.
5.1 Replacing the Fuse
CAUTION: REMOVE POWER TO UNIT BEFORE REPLACING THE FUSE.
When a fuse blows, check first to determine the cause, then replace the fuse using the following procedure:
1.Disconnect the AC power and place the power switch located on the rear panel in the O position. Remove the power cord from the receptacle.
2.The fuse receptacle is located in the power cord receptacle assembly in the upper
Figure 5-1: AC Fuse Replacement