Teledyne 9060L, Percent Oxygen Analyzer Electrical Connections, Primary Input Power

Page 25

Percent Oxygen Analyzer

Installation

 

 

 

probe or install a longer probe. This is normally caused by stratification of the flue gas.

Figure 3-1: Oxygen Probe Mounting

3.4 Electrical Connections

Figure 3-1 shows the Model 9060L rear panel. For detailed pinouts, see the wiring/interconnection drawings in the Drawings section at the rear of this manual.

3.4.1 Primary Input Power

The power cord receptacle, fuse block and Power switch are located in the same assembly. A 6-foot, standard AC power cord is supplied with the control unit. Insert the female plug end of the power cord into the power cord receptacle.

The power supply allows direct connection to any 110 VAC, 50/60Hz power source. 220VAC is an optional configuration settable through an internal switch in the control unit. The fuse block, to the right of the power cord receptacle, accepts two 5x20mm IEC type T, 2 A, time-lag (T) fuse. (See Fuse Replacement in chapter 5, Maintenance.)

The Power switch is located on the right-hand end of the power source input receptacle assembly.

Teledyne Analytical Instruments

25

Image 25
Contents Teledyne Analytical Instruments Model 9060LCopyright 2005 Teledyne Instruments/ Analytical Instruments Model 9060LPercent Oxygen Analyzer Specific Model InformationSafety Messages Percent Oxygen Analyzer Table of Contents Operation MaintenanceAppendix Index Teledyne Analytical InstrumentsFront Panel List of FiguresList of Tables Combustible GAS Usage Introduction Main Features of the AnalyzerOverview Percent Oxygen Analyzer IntroductionIntroduction Model 9060L Front Panel DescriptionSpan Rear Panel DescriptionRear Panel Introduction Operational TheoryOxygen Probe, Sensor, and Heater Percent Oxygen Analyzer Operational TheoryModel 9060L Oxygen Probe Operational Theory Model 9060LProbe Options Probe ConfigurationInternal Structure of the Zirconia Sensor and Probe Zirconia SensorHeater Sensor Impedance Applications Not at Atmospheric PressureElectronics GeneralSignal Processing Block Diagram Signal ProcessingOperational Theory Model 9060L Control Unit Installation Installation9060L Oxygen Probe Installation Percent Oxygen Analyzer InstallationInstallation Model 9060L Electrical Connections Primary Input PowerAnalog Outputs Alarm RelaysAlarm #2 4Model 9060L Heated Probe Cable Connection Oxygen Probe ConnectionsConnections to Probe Installation ChecklistInstallation Model 9060L Teledyne Analytical Instruments Percent Oxygen Analyzer Operation OperationSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 9060L Set Alarm 1 HI alarm Setting the Alarm SetpointsHI Range LO RangeSensor Fail Alarm Set Alarm 2 LO alarmSelecting a Fixed Range or Autoranging Span Calibration Oxygen CalibrationOperation Model 9060L Supplementary Information Probe TemperatureAir Calibration Low Temperature Error MessagesOpen Thermocouple Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Percent Oxygen Analyzer Maintenance MaintenanceReplacing the Fuse Maintenance Model 9060L Appendix SpecificationsAppendix Model 9060L Probe SpecificationsEmail asktai@teledyne.com Spare Parts ListAnalytical Instruments Probe or Sensor EMF Tables Reference DrawingZirconia Oxygen Sensor Output mV Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer