Installation

Model 9060L

 

 

 

Figure 3-2: Model 9060L Rear panel

3.4.2 Analog Outputs

There are three DC output signal connectors with screw terminals on the panel. There are two wires per output with the polarity noted. See Figure 3-3. The outputs are:

0–10V % Range:

Voltage rises with increasing oxygen

 

concentration, from 0V at 0 percent oxygen to

 

10V at full scale percent oxygen. (Full scale =

 

100% of programmed range.)

0–10V Range ID:

03.33V = Low Range, 06.66V = High Range,

 

10V = Air Cal Range.

4–20 mA % Range:

Current increases with increasing oxygen

 

concentration, from 4 mA at 0 percent oxygen to

 

20 mA at full scale percent oxygen. (Full scale =

 

100% of programmed range.)

3.4.3 Alarm Relays

The three alarm-circuit connectors are screw terminals for making connections to internal alarm relay contacts. There is one set of contacts for each type of alarm. Contacts are Form C, with normally open and normally closed contact connections capable of switching up to 1.0 ampere at 125 VAC into a resistive load (2A for 30 VDC).

Teledyne Analytical Instruments

26

Image 26
Contents Model 9060L Teledyne Analytical InstrumentsModel 9060L Copyright 2005 Teledyne Instruments/ Analytical InstrumentsSpecific Model Information Percent Oxygen AnalyzerSafety Messages Percent Oxygen Analyzer Table of Contents Appendix MaintenanceOperation Index Teledyne Analytical InstrumentsList of Figures Front PanelList of Tables Combustible GAS Usage Overview Main Features of the AnalyzerIntroduction Percent Oxygen Analyzer IntroductionFront Panel Description Introduction Model 9060LRear Panel Description SpanRear Panel Oxygen Probe, Sensor, and Heater Operational TheoryIntroduction Percent Oxygen Analyzer Operational TheoryOperational Theory Model 9060L Model 9060L Oxygen ProbeProbe Configuration Probe OptionsZirconia Sensor Internal Structure of the Zirconia Sensor and ProbeHeater Electronics Applications Not at Atmospheric PressureSensor Impedance GeneralSignal Processing Signal Processing Block DiagramOperational Theory Model 9060L 9060L Oxygen Probe Installation InstallationControl Unit Installation Percent Oxygen Analyzer InstallationInstallation Model 9060L Primary Input Power Electrical ConnectionsAlarm Relays Analog OutputsAlarm #2 Oxygen Probe Connections 4Model 9060L Heated Probe Cable ConnectionInstallation Checklist Connections to ProbeInstallation Model 9060L Teledyne Analytical Instruments Operation Percent Oxygen Analyzer OperationOperation Model 9060L Using the Function and Data Entry ButtonsSetting the Analysis Ranges HI Range Setting the Alarm SetpointsSet Alarm 1 HI alarm LO RangeSelecting a Fixed Range or Autoranging Set Alarm 2 LO alarmSensor Fail Alarm Oxygen Calibration Span CalibrationOperation Model 9060L Air Calibration Probe TemperatureSupplementary Information Open Thermocouple Error MessagesLow Temperature Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Replacing the Fuse MaintenancePercent Oxygen Analyzer Maintenance Maintenance Model 9060L Specifications AppendixProbe Specifications Appendix Model 9060LAnalytical Instruments Spare Parts ListEmail asktai@teledyne.com Zirconia Oxygen Sensor Output mV Reference DrawingProbe or Sensor EMF Tables Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer