Teledyne 9060L, Percent Oxygen Analyzer operating instructions Heater

Page 19

Percent Oxygen Analyzer

Operational Theory

 

 

 

require a heater. When the outside and inside of the sensor are exposed to different oxygen partial pressures, an EMF (E) is developed which obeys the Nernst equation:

 

RT

(PO2) Inside

E(millivolts) =

 

log e

 

4F

 

 

(PO2) Outside

where T is the temperature (K) at the disc (>650°C (1200°F)), R is the gas constant. F is the Faraday constant and (PO2) Inside and (PO2) Outside are the oxygen partial pressures at the inner and outer electrodes respectively with the higher oxygen partial pressure being positive.

If dry air at atmospheric pressure, (21% oxygen) is used as a reference at the inner electrode, the following equations are obtained:

 

− 2

0.21

E(millivolts) = 2.154 ⋅10

 

T loge

 

 

 

 

 

(PO2 ) Outside

Transposing this equation:

 

 

⎛ - 46.421E

(%O2 ) Outside (ATM) = 0.21 EXP ⎜

 

T

The 9060L transmitter solves this equation which is valid above 650°C (1200°F). Either the probe heater or the process maintains the sensor temperature at this level.

2.2.3 Heater

CAUTION: DANGEROUS VOLTAGE (120 VAC) EXISTS WITHIN THE PROBE OR HEATER AND PRESENTS AN ELECTRICAL SHOCK HAZARD TO MAINTENANCE PERSONNEL. ALWAYS DISCONNECT THE PROBE FROM THE ANALYZER BEFORE WORKING WITH THE PROBE, SENSOR, OR HEATER.

THE EARTH CONNECTION (GREEN WIRE) FROM

Teledyne Analytical Instruments

19

Image 19
Contents Teledyne Analytical Instruments Model 9060LCopyright 2005 Teledyne Instruments/ Analytical Instruments Model 9060LPercent Oxygen Analyzer Specific Model InformationSafety Messages Percent Oxygen Analyzer Table of Contents Index Teledyne Analytical Instruments MaintenanceOperation AppendixFront Panel List of FiguresList of Tables Combustible GAS Usage Percent Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewIntroduction Model 9060L Front Panel DescriptionSpan Rear Panel DescriptionRear Panel Percent Oxygen Analyzer Operational Theory Operational TheoryIntroduction Oxygen Probe, Sensor, and HeaterModel 9060L Oxygen Probe Operational Theory Model 9060LProbe Options Probe ConfigurationInternal Structure of the Zirconia Sensor and Probe Zirconia SensorHeater General Applications Not at Atmospheric PressureSensor Impedance ElectronicsSignal Processing Block Diagram Signal ProcessingOperational Theory Model 9060L Percent Oxygen Analyzer Installation InstallationControl Unit Installation 9060L Oxygen Probe InstallationInstallation Model 9060L Electrical Connections Primary Input PowerAnalog Outputs Alarm RelaysAlarm #2 4Model 9060L Heated Probe Cable Connection Oxygen Probe ConnectionsConnections to Probe Installation ChecklistInstallation Model 9060L Teledyne Analytical Instruments Percent Oxygen Analyzer Operation OperationSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 9060L LO Range Setting the Alarm SetpointsSet Alarm 1 HI alarm HI RangeSensor Fail Alarm Set Alarm 2 LO alarmSelecting a Fixed Range or Autoranging Span Calibration Oxygen CalibrationOperation Model 9060L Supplementary Information Probe TemperatureAir Calibration Low Temperature Error MessagesOpen Thermocouple Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Percent Oxygen Analyzer Maintenance MaintenanceReplacing the Fuse Maintenance Model 9060L Appendix SpecificationsAppendix Model 9060L Probe SpecificationsEmail asktai@teledyne.com Spare Parts ListAnalytical Instruments Probe or Sensor EMF Tables Reference DrawingZirconia Oxygen Sensor Output mV Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer