Teledyne Percent Oxygen Analyzer operating instructions Operation Model 9060L

Page 36

Operation

Model 9060L

 

 

 

for five minutes before timing-out. During this five minute interval, the LED will continue to track gas through the analyzer.

2.Flow span gas of a known oxygen concentration (between 1 and 16% oxygen) into the analyzer. The analyzer will begin to track the span gas.

Note: The alarms will not change state during span calibration.

3.When the meter stabilizes, use the UP/DOWN arrows to adjust the analyzer span value to the exact percent O2 concentration in the span gas. Remember that the acceptable range is 1 to 16% of adjustment. The display will freeze at 1% or 16% if you try to adjust beyond those limits.

Note: When an arrow button is first pressed, the LED begins flashing slightly more rapidly and no longer tracks the span gas. Instead, it responds to the UP/DOWN keystrokes.

Note: While the LED is flashing slightly more rapidly, the SPAN

routine will time-out in five seconds (instead of five minutes), if no further keystrokes are entered.

4.When the span value is set to the known concentration of O2 in the span gas, stop pressing the keys and wait for five seconds.

Note: When you stop pressing the UP/DOWN keys, the rapid flashing will cease and the analyzer will acquire the new span value after five seconds.

The alarms will only be non-responsive for 60 seconds. This time frame allows you to reintroduce sample gas into the analyzer.

5.Shut off the span gas flow and immediately flow sample into the analyzer.

6.After a few minutes, disconnect the calibration gas line and replace the plug to cap off the calibration port.

Teledyne Analytical Instruments

36

Image 36
Contents Model 9060L Teledyne Analytical InstrumentsModel 9060L Copyright 2005 Teledyne Instruments/ Analytical InstrumentsSpecific Model Information Percent Oxygen AnalyzerSafety Messages Percent Oxygen Analyzer Table of Contents Maintenance OperationAppendix Index Teledyne Analytical InstrumentsList of Figures Front PanelList of Tables Combustible GAS Usage Main Features of the Analyzer IntroductionOverview Percent Oxygen Analyzer IntroductionFront Panel Description Introduction Model 9060LRear Panel Description SpanRear Panel Operational Theory IntroductionOxygen Probe, Sensor, and Heater Percent Oxygen Analyzer Operational TheoryOperational Theory Model 9060L Model 9060L Oxygen ProbeProbe Configuration Probe OptionsZirconia Sensor Internal Structure of the Zirconia Sensor and ProbeHeater Applications Not at Atmospheric Pressure Sensor ImpedanceElectronics GeneralSignal Processing Signal Processing Block DiagramOperational Theory Model 9060L Installation Control Unit Installation9060L Oxygen Probe Installation Percent Oxygen Analyzer InstallationInstallation Model 9060L Primary Input Power Electrical ConnectionsAlarm Relays Analog OutputsAlarm #2 Oxygen Probe Connections 4Model 9060L Heated Probe Cable ConnectionInstallation Checklist Connections to ProbeInstallation Model 9060L Teledyne Analytical Instruments Operation Percent Oxygen Analyzer OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation Model 9060L Setting the Alarm Setpoints Set Alarm 1 HI alarmHI Range LO RangeSet Alarm 2 LO alarm Sensor Fail AlarmSelecting a Fixed Range or Autoranging Oxygen Calibration Span CalibrationOperation Model 9060L Probe Temperature Supplementary InformationAir Calibration Error Messages Low TemperatureOpen Thermocouple Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Maintenance Percent Oxygen Analyzer MaintenanceReplacing the Fuse Maintenance Model 9060L Specifications AppendixProbe Specifications Appendix Model 9060LSpare Parts List Email asktai@teledyne.comAnalytical Instruments Reference Drawing Probe or Sensor EMF TablesZirconia Oxygen Sensor Output mV Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer