Percent Oxygen Analyzer

Appendix

 

 

 

11.0

13.78

10.5

14.78

10.0

15.82

9.5

16.92

9.0

18.08

8.5

19.30

8.0

20.60

7.5

21.98

7.0

23.45

6.5

25.04

6.0

26.75

5.5

28.61

5.0

30.65

4.5

32.90

4.0

35.42

3.5

38.28

3.0

41.58

2.5

45.48

2.0

50.25

1.5

56.41

1.0

65.08

0.5

79.91

0.2

99.51

'K’ TC mV 29.212 at 720 °C (1320°F)

These tables are based on the Nernst equation:

Sensor e.m.f. = 0.02154 x T x 1n ( 20.95 / % oxygen),

where T = ° K (° C + 273), e.m.f. is in mV’s

Teledyne Analytical Instruments

47

Image 47
Contents Teledyne Analytical Instruments Model 9060LCopyright 2005 Teledyne Instruments/ Analytical Instruments Model 9060LPercent Oxygen Analyzer Specific Model InformationSafety Messages Percent Oxygen Analyzer Table of Contents Index Teledyne Analytical Instruments MaintenanceOperation AppendixFront Panel List of FiguresList of Tables Combustible GAS Usage Percent Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewIntroduction Model 9060L Front Panel DescriptionSpan Rear Panel DescriptionRear Panel Percent Oxygen Analyzer Operational Theory Operational TheoryIntroduction Oxygen Probe, Sensor, and HeaterModel 9060L Oxygen Probe Operational Theory Model 9060LProbe Options Probe ConfigurationInternal Structure of the Zirconia Sensor and Probe Zirconia SensorHeater General Applications Not at Atmospheric PressureSensor Impedance ElectronicsSignal Processing Block Diagram Signal ProcessingOperational Theory Model 9060L Percent Oxygen Analyzer Installation InstallationControl Unit Installation 9060L Oxygen Probe InstallationInstallation Model 9060L Electrical Connections Primary Input PowerAnalog Outputs Alarm RelaysAlarm #2 4Model 9060L Heated Probe Cable Connection Oxygen Probe ConnectionsConnections to Probe Installation ChecklistInstallation Model 9060L Teledyne Analytical Instruments Percent Oxygen Analyzer Operation OperationOperation Model 9060L Using the Function and Data Entry ButtonsSetting the Analysis Ranges LO Range Setting the Alarm SetpointsSet Alarm 1 HI alarm HI RangeSelecting a Fixed Range or Autoranging Set Alarm 2 LO alarmSensor Fail Alarm Span Calibration Oxygen CalibrationOperation Model 9060L Air Calibration Probe TemperatureSupplementary Information Open Thermocouple Error MessagesLow Temperature Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Replacing the Fuse MaintenancePercent Oxygen Analyzer Maintenance Maintenance Model 9060L Appendix SpecificationsAppendix Model 9060L Probe SpecificationsAnalytical Instruments Spare Parts ListEmail asktai@teledyne.com Zirconia Oxygen Sensor Output mV Reference DrawingProbe or Sensor EMF Tables Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer