Teledyne 9060L, Percent Oxygen Analyzer operating instructions Signal Processing Block Diagram

Page 21

Percent Oxygen Analyzer

Operational Theory

 

 

 

located on the Main PCB, which is mounted vertically, just behind the front panel of the control unit.

2.5.2 Signal Processing

Figure 2-3 is a block diagram of the signal processing electronics described below.

Figure 2-3: Signal Processing Block Diagram

In the presence of oxygen the sensor generates a millivolt potential which is fed to a voltage amplifier. The output from the sensor amplifier is sent to an analog to digital converter (ADC), and the resulting digital concentration signal is passed to the micro-controller. The ADC has a second channel which monitors the temperature of the probe.

Teledyne Analytical Instruments

21

Image 21 Contents
Teledyne Analytical Instruments Model 9060LCopyright 2005 Teledyne Instruments/ Analytical Instruments Model 9060LPercent Oxygen Analyzer Specific Model InformationSafety Messages Percent Oxygen Analyzer Table of Contents Operation MaintenanceAppendix Index Teledyne Analytical InstrumentsFront Panel List of FiguresList of Tables Combustible GAS Usage Introduction Main Features of the AnalyzerOverview Percent Oxygen Analyzer IntroductionIntroduction Model 9060L Front Panel DescriptionSpan Rear Panel DescriptionRear Panel Introduction Operational TheoryOxygen Probe, Sensor, and Heater Percent Oxygen Analyzer Operational TheoryModel 9060L Oxygen Probe Operational Theory Model 9060LProbe Options Probe ConfigurationInternal Structure of the Zirconia Sensor and Probe Zirconia SensorHeater Sensor Impedance Applications Not at Atmospheric PressureElectronics GeneralSignal Processing Block Diagram Signal ProcessingOperational Theory Model 9060L Control Unit Installation Installation9060L Oxygen Probe Installation Percent Oxygen Analyzer InstallationInstallation Model 9060L Electrical Connections Primary Input PowerAnalog Outputs Alarm RelaysAlarm #2 4Model 9060L Heated Probe Cable Connection Oxygen Probe ConnectionsConnections to Probe Installation ChecklistInstallation Model 9060L Teledyne Analytical Instruments Percent Oxygen Analyzer Operation OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation Model 9060L Set Alarm 1 HI alarm Setting the Alarm SetpointsHI Range LO RangeSet Alarm 2 LO alarm Sensor Fail AlarmSelecting a Fixed Range or Autoranging Span Calibration Oxygen CalibrationOperation Model 9060L Probe Temperature Supplementary InformationAir Calibration Error Messages Low TemperatureOpen Thermocouple Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Maintenance Percent Oxygen Analyzer MaintenanceReplacing the Fuse Maintenance Model 9060L Appendix SpecificationsAppendix Model 9060L Probe SpecificationsSpare Parts List Email asktai@teledyne.comAnalytical Instruments Reference Drawing Probe or Sensor EMF TablesZirconia Oxygen Sensor Output mV Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer
Related manuals
Manual 51 pages 50.18 Kb