Teledyne 9060L Supplementary Information, Probe Temperature, Air Calibration

Page 37

Percent Oxygen Analyzer

Operation

 

 

 

4.6.2 Air Calibration

To perform air calibration, make sure you have air flowing through the same inlet you use for span calibration at a flow of 1- 5 liters per minute. Wait for readings to stabilize.

Press both the SET ALARM 1 and SET ALARM 2 simultaneously. The LED display will show “OFSt” for a few seconds. After that the instrument should read near 20.9%.

The sensor output is expected to be near zero millivolts, if the offset of the sensor is too high, the analyzer will not calibrate.

4.7 Supplementary Information

If, during the span procedure, you pressed the SPAN button by mistake, you must wait five minutes for the analyzer to resume analysis or you can press the UP button and then the DOWN button. (Pressing the UP and DOWN buttons causes the analyzer to time-out in five seconds instead of five minutes).

If during the span procedure, you press the RANGE or ALARM buttons:

-either the range or alarm routine will be activated.

-any changes to span will be rejected.

-the 60 second alarm delay will not occur, i.e., the alarms will be responsive immediately.

CAUTION: TI/AI CONSIDERS THE ACTION OF PRESSING THE ALARM OR RANGE BUTTONS TO BE AT YOUR (THE USER’S) DISCRETION AND NOT UNDER GUARANTEE OF ALARM PROTECTION.

4.8 Probe Temperature

The probe temperature can be displayed momentarily and adjusted if needed. While the instrument is in normal mode, displaying oxygen concentration, press the UP and DOWN arrow keys simultaneously. The red LED display will show the probe temperature in degrees centigrade for five seconds.

Teledyne Analytical Instruments

37

Image 37
Contents Teledyne Analytical Instruments Model 9060LCopyright 2005 Teledyne Instruments/ Analytical Instruments Model 9060LPercent Oxygen Analyzer Specific Model InformationSafety Messages Percent Oxygen Analyzer Table of Contents Operation MaintenanceAppendix Index Teledyne Analytical InstrumentsFront Panel List of FiguresList of Tables Combustible GAS Usage Introduction Main Features of the AnalyzerOverview Percent Oxygen Analyzer IntroductionIntroduction Model 9060L Front Panel DescriptionSpan Rear Panel DescriptionRear Panel Introduction Operational TheoryOxygen Probe, Sensor, and Heater Percent Oxygen Analyzer Operational TheoryModel 9060L Oxygen Probe Operational Theory Model 9060LProbe Options Probe ConfigurationInternal Structure of the Zirconia Sensor and Probe Zirconia SensorHeater Sensor Impedance Applications Not at Atmospheric PressureElectronics GeneralSignal Processing Block Diagram Signal ProcessingOperational Theory Model 9060L Control Unit Installation Installation9060L Oxygen Probe Installation Percent Oxygen Analyzer InstallationInstallation Model 9060L Electrical Connections Primary Input PowerAnalog Outputs Alarm RelaysAlarm #2 4Model 9060L Heated Probe Cable Connection Oxygen Probe ConnectionsConnections to Probe Installation ChecklistInstallation Model 9060L Teledyne Analytical Instruments Percent Oxygen Analyzer Operation OperationSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 9060L Set Alarm 1 HI alarm Setting the Alarm SetpointsHI Range LO RangeSensor Fail Alarm Set Alarm 2 LO alarmSelecting a Fixed Range or Autoranging Span Calibration Oxygen CalibrationOperation Model 9060L Supplementary Information Probe TemperatureAir Calibration Low Temperature Error MessagesOpen Thermocouple Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Percent Oxygen Analyzer Maintenance MaintenanceReplacing the Fuse Maintenance Model 9060L Appendix SpecificationsAppendix Model 9060L Probe SpecificationsEmail asktai@teledyne.com Spare Parts ListAnalytical Instruments Probe or Sensor EMF Tables Reference DrawingZirconia Oxygen Sensor Output mV Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer