Teledyne 9060L, Percent Oxygen Analyzer operating instructions Alarm #2

Page 27

Percent Oxygen Analyzer

Installation

 

 

 

The alarm relay circuits are designed for failsafe operation, meaning the relays are energized during normal operation. If power fails the relays de-energize (alarms activated).

The contact connections are indicated diagrammatically on the rear panel as Normally Closed, Common, and Normally Open. Figure 3-3 explains how these act in failsafe operation.

Alarm 1 and Alarm 2 are configured as HI and LO respectively. A HI alarm will activate when concentration is above threshold, while a LO alarm will activate when concentration is below threshold.

Figure 3-3: Contact ID for FAILSAFE Relay Operation

The specific descriptions for each type of alarm are as follows:

Alarm #1

Programmable as high alarm. Can be set anywhere

 

between 0 and 100 % but no lower than Alarm #2.

Alarm #2

Programmable as low alarm. Can be set anywhere

 

from 0 to 100 % but no higher than Alarm #1.

Sensor Fail

Actuates when the output of the sensor falls below a

 

factory programmed acceptable level or the

 

thermocouple is open.

CAUTION:

THERE COULD BE HAZARDOUS VOLTAGE AT THE

 

ALARMS TERMINALS, EVEN WHEN POWER IS

 

REMOVED FROM THE INSTRUMENT.

Teledyne Analytical Instruments

27

Image 27
Contents Teledyne Analytical Instruments Model 9060LCopyright 2005 Teledyne Instruments/ Analytical Instruments Model 9060LPercent Oxygen Analyzer Specific Model InformationSafety Messages Percent Oxygen Analyzer Table of Contents Index Teledyne Analytical Instruments MaintenanceOperation AppendixFront Panel List of FiguresList of Tables Combustible GAS Usage Percent Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewIntroduction Model 9060L Front Panel DescriptionSpan Rear Panel DescriptionRear Panel Percent Oxygen Analyzer Operational Theory Operational TheoryIntroduction Oxygen Probe, Sensor, and HeaterModel 9060L Oxygen Probe Operational Theory Model 9060LProbe Options Probe ConfigurationInternal Structure of the Zirconia Sensor and Probe Zirconia SensorHeater General Applications Not at Atmospheric PressureSensor Impedance ElectronicsSignal Processing Block Diagram Signal ProcessingOperational Theory Model 9060L Percent Oxygen Analyzer Installation InstallationControl Unit Installation 9060L Oxygen Probe InstallationInstallation Model 9060L Electrical Connections Primary Input PowerAnalog Outputs Alarm RelaysAlarm #2 4Model 9060L Heated Probe Cable Connection Oxygen Probe ConnectionsConnections to Probe Installation ChecklistInstallation Model 9060L Teledyne Analytical Instruments Percent Oxygen Analyzer Operation OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation Model 9060L LO Range Setting the Alarm SetpointsSet Alarm 1 HI alarm HI RangeSet Alarm 2 LO alarm Sensor Fail AlarmSelecting a Fixed Range or Autoranging Span Calibration Oxygen CalibrationOperation Model 9060L Probe Temperature Supplementary InformationAir Calibration Error Messages Low TemperatureOpen Thermocouple Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Maintenance Percent Oxygen Analyzer MaintenanceReplacing the Fuse Maintenance Model 9060L Appendix SpecificationsAppendix Model 9060L Probe SpecificationsSpare Parts List Email asktai@teledyne.comAnalytical Instruments Reference Drawing Probe or Sensor EMF TablesZirconia Oxygen Sensor Output mV Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer