Teledyne Percent Oxygen Analyzer Using the Function and Data Entry Buttons, Operation Model 9060L

Page 32

Operation

Model 9060L

 

 

 

4.2 Using the Function and Data Entry Buttons

When no buttons on the analyzer are being pressed, the instrument is in the analyze mode. It is monitoring the percent of oxygen in the sample gas that is flowing through the remote probe.

When one of the function buttons is being pressed, the analyzer enters the setup mode or the calibration mode.

Setup Function Buttons:

There are 4 setup function buttons on the analyze:

SET ALARM 1

SET ALARM 2

SET HI RANGE

SET LO RANGE

Calibration Button:

The calibration mode button is:

SPAN

Data Entry buttons:

The data entry buttons (Δ and ∇) increment the values displayed on the PERCENT OXYGEN meter while one of the function buttons is being held down.

Δ : Increments the displayed value upwards.

∇ : Increments the displayed value downwards.

Any of the functions can be selected at any time by holding down the appropriate button.

Each function will be described in the following sections. Although the operator can use any function at any time, the order chosen in this manual is appropriate for an initial setup.

4.3 Setting the Analysis Ranges

The two user definable analysis ranges are both capable of being adjusted for from 0-1 to 0-100% oxygen concentration. Whatever values are selected, the analyzer automatically switches from the LO range to the HI range when the oxygen concentration reaches 100% of the LO

Teledyne Analytical Instruments

32

Image 32
Contents Model 9060L Teledyne Analytical InstrumentsModel 9060L Copyright 2005 Teledyne Instruments/ Analytical InstrumentsSpecific Model Information Percent Oxygen AnalyzerSafety Messages Percent Oxygen Analyzer Table of Contents Maintenance OperationAppendix Index Teledyne Analytical InstrumentsList of Figures Front PanelList of Tables Combustible GAS Usage Main Features of the Analyzer IntroductionOverview Percent Oxygen Analyzer IntroductionFront Panel Description Introduction Model 9060LRear Panel Description SpanRear Panel Operational Theory IntroductionOxygen Probe, Sensor, and Heater Percent Oxygen Analyzer Operational TheoryOperational Theory Model 9060L Model 9060L Oxygen ProbeProbe Configuration Probe OptionsZirconia Sensor Internal Structure of the Zirconia Sensor and ProbeHeater Applications Not at Atmospheric Pressure Sensor ImpedanceElectronics GeneralSignal Processing Signal Processing Block DiagramOperational Theory Model 9060L Installation Control Unit Installation9060L Oxygen Probe Installation Percent Oxygen Analyzer InstallationInstallation Model 9060L Primary Input Power Electrical ConnectionsAlarm Relays Analog OutputsAlarm #2 Oxygen Probe Connections 4Model 9060L Heated Probe Cable ConnectionInstallation Checklist Connections to ProbeInstallation Model 9060L Teledyne Analytical Instruments Operation Percent Oxygen Analyzer OperationOperation Model 9060L Using the Function and Data Entry ButtonsSetting the Analysis Ranges Setting the Alarm Setpoints Set Alarm 1 HI alarmHI Range LO RangeSelecting a Fixed Range or Autoranging Set Alarm 2 LO alarmSensor Fail Alarm Oxygen Calibration Span CalibrationOperation Model 9060L Air Calibration Probe TemperatureSupplementary Information Open Thermocouple Error MessagesLow Temperature Percent Oxygen Analyzer Operation Operation Model 9060L Teledyne Analytical Instruments Replacing the Fuse MaintenancePercent Oxygen Analyzer Maintenance Maintenance Model 9060L Specifications AppendixProbe Specifications Appendix Model 9060LAnalytical Instruments Spare Parts ListEmail asktai@teledyne.com Zirconia Oxygen Sensor Output mV Reference DrawingProbe or Sensor EMF Tables Percent Oxygen Analyzer Appendix Model 9060L Teledyne Analytical Instruments Index Index Model 9060L Percent Oxygen Analyzer