Installation

BDS 3960

 

 

 

 

 

 

3.3.2.1PRIMARY INPUT POWER

The power cord receptacle and fuse block are located in the same assembly. Insert the power cord into the power cord receptacle.

CAUTION: POWER IS APPLIED TO THE INSTRUMENT'S CIRCUITRY AS LONG AS THE INSTRUMENT IS CONNECTED TO THE POWER SOURCE.

The power supply requires 100–120 VAC, 50/60 Hz power source.

Fuse Installation: The fuse block, at the right of the power cord receptacle, accepts US or European size fuses. A jumper replaces the fuse in whichever fuse receptacle is not used. Fuses may not be installed at the factory. Be sure to install the proper fuse as part of installation. (See Fuse Replacement in Chapter 5, Maintenance.)

3.3.2.250-PIN EQUIPMENT INTERFACE CONNECTOR

Figure 3-4 shows the pin layout of the Equipment Interface connector. The arrangement is shown as seen when the viewer faces the rear panel of the analyzer. The pin numbers for each input/output function are given where each function is described in the paragraphs below.

Figure 3-4: Equipment Interface Connector Pin Arrangement

Analog Outputs: There are four DC output signal pins—two pins per output. For polarity, see Table 3-1. The outputs are:

0–1 VDC % of Range:

Voltage rises linearly with increasing

 

oxygen, from 0 V at 0 ppm to 1 V at

 

full scale ppm. (Full scale = 100% of

 

programmable range.)

0–1 VDC Range ID:

0.25 V = Low Range, 0.5 V = Medium

 

Range, 0.75 V = High Range, 1 V =

 

100ppm.

Teledyne Analytical Instruments

34

Image 34
Contents Oxygen Analyzer Teledyne Analytical InstrumentsBDS Copyright 2002 Teledyne Analytical InstrumentsSpecific Model Information IiiSafety Messages Oxygen Analyzer Table of Contents Operation ViiIndex Teledyne Analytical Instruments MaintenanceAppendix ViiiList of Figures BDS-3960 Front PanelVacuum Degassing for the BDS Oxygen Sensor List of Tables Combustible GAS Usage XiiOverview Main Features of the AnalyzerIntroduction Typical ApplicationsFront Panel Operator Interface BDS-3960 Front Panel Remote Probe Rear Panel Equipment InterfaceAnalog Outputs Range ID ContactsModel BDS 3960 Rear Panel BDS Sensor Operational TheoryIntroduction Principles of OperationOxygen Analyzer Operational Theory Cross Section of the BDS Oxygen Sensor Gas Flow Rate BDS sensor output at different gas flow rateRecovery from High Level Oxygen Exposure Temperature effectGas Pressure Background Gas Compatibility Typical Purge-down Curve After Air SaturationMaintenance StabilitySample System Electronics and Signal Processing Flow DiagramOxygen Analyzer Operational Theory BDS 3960 Electronics Block Diagram Page Unpacking the Analyzer InstallationMounting the Analyzer Oxygen Analyzer Installation Model BDS 3960 Front PanelRear Panel Connections Gas ConnectionsElectrical Connections Equipment Interface Connector Pin Arrangement Threshold Alarm Pin FunctionSystem Alarm Pin ContactCal Contact Installation Electrolyte Refill of BDS Sensor Testing the System Adding Electrolyte to the BDS SensorPowering Up the System Operation Auto Ranging onOxygen Analyzer Maintenance Analyzer applicationSystem Screen Communication Information and Calibration ParametersSetting Software Parameters Filter, Gas Factor, Tmp. Coeff Gas Factor Parameter Selection BoxTemperature Coefficient Parameter Default Values Calibration of the Analyzer Zero CalSpan Cal Alarms Function Maintenance Setting the Analog Output Ranges Range FunctionRange Options List Box Signal Output Fixed Range AnalysisSwitching The Program Back To Front Oxygen Analyzer Maintenance Maintenance Routine Maintenance Adding Water to the BDS SensorFuse Replacement Adding Water into the BDS sensorBattery Backup Replacement Removing Fuse Block from HousingReinstalling Application software to PPC Click startProgramsexplorer Click showMy DeviceSymptoms Possible causes and Solutions CleaningTroubleshooting Major Internal ComponentsCauses and solution keys Vacuum Degassing for the BDS Oxygen Sensor Oxygen Analyzer Maintenance Teledyne Analytical Instruments Specifications AppendixOperating Temperature 5-40 C Oxygen Analyzer AppendixRecommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountApplication notes Section I Product Identification Material Safety Data SheetSection II Physical and Chemical Data Section III -Physical Hazards Section IV Health Hazard DataSection VI Handling Information Alarms IndexOxygen Analyzer Index MsdsZero