2. Are either or both of the alarms to be configured as failsafe?

In failsafe mode, the alarm relay de-energizes in an alarm condition. For non-failsafe operation, the relay is energized in an alarm condition. You can set either or both of the concentration alarms to operate in failsafe or non-failsafe mode.

3. Are either of the alarms to be latching?

In latching mode, once the alarm or alarms trigger, they will remain in the alarm mode even if process conditions revert back to non-alarm conditions. This mode requires an alarm to be recognized before it can be reset. In the non-latching mode, the alarm status will terminate when process conditions revert to non-alarm conditions.

4. Are either of the alarms to be defeated?

The defeat alarm mode is incorporated into the alarm circuit so that maintenance can be performed under conditions which would normally activate the alarms.

The defeat function can also be used to reset a latched alarm. (See procedures, below.)

Each of the concentration alarms have their own button on Main display of the PPC application, ALARM 1 and ALARM 2. Once you press on the alarm buttons you will be presented with alarm setup screen.

Five parameters can be changed on the alarm setup screen:

1. Value of the alarm level—#### ppb (oxygen); value can be set from 0 to 100,000 ppb.

2.Out-of-range direction—High or Low

3.Latch control—Latching or Non-latching.

4.On/off Control—Active or Defeated.

5.Safety Mode—Failsafe or Non-failsafe.

To reset a latched alarm, go to On/off Control and select defeated.

Teledyne Analytical Instruments

52

Image 52
Contents Oxygen Analyzer Teledyne Analytical InstrumentsBDS Copyright 2002 Teledyne Analytical InstrumentsSpecific Model Information IiiSafety Messages Oxygen Analyzer Table of Contents Operation ViiMaintenance AppendixIndex Teledyne Analytical Instruments ViiiList of Figures BDS-3960 Front PanelVacuum Degassing for the BDS Oxygen Sensor List of Tables Combustible GAS Usage XiiMain Features of the Analyzer IntroductionOverview Typical ApplicationsFront Panel Operator Interface BDS-3960 Front Panel Rear Panel Equipment Interface Analog OutputsRemote Probe Range ID ContactsModel BDS 3960 Rear Panel Operational Theory IntroductionBDS Sensor Principles of OperationOxygen Analyzer Operational Theory Cross Section of the BDS Oxygen Sensor Gas Flow Rate BDS sensor output at different gas flow rateRecovery from High Level Oxygen Exposure Temperature effectGas Pressure Background Gas Compatibility Typical Purge-down Curve After Air SaturationMaintenance StabilitySample System Electronics and Signal Processing Flow DiagramOxygen Analyzer Operational Theory BDS 3960 Electronics Block Diagram Page Unpacking the Analyzer InstallationMounting the Analyzer Oxygen Analyzer Installation Model BDS 3960 Front PanelRear Panel Connections Gas ConnectionsElectrical Connections Equipment Interface Connector Pin Arrangement Threshold Alarm Pin FunctionSystem Alarm Pin ContactCal Contact Installation Electrolyte Refill of BDS Sensor Testing the System Adding Electrolyte to the BDS SensorPowering Up the System Operation Auto Ranging onOxygen Analyzer Maintenance Analyzer applicationSystem Screen Communication Information and Calibration ParametersSetting Software Parameters Filter, Gas Factor, Tmp. Coeff Gas Factor Parameter Selection BoxTemperature Coefficient Parameter Default Values Calibration of the Analyzer Zero CalSpan Cal Alarms Function Maintenance Setting the Analog Output Ranges Range FunctionRange Options List Box Signal Output Fixed Range AnalysisSwitching The Program Back To Front Oxygen Analyzer Maintenance Maintenance Routine Maintenance Adding Water to the BDS SensorFuse Replacement Adding Water into the BDS sensorBattery Backup Replacement Removing Fuse Block from HousingReinstalling Application software to PPC Click startProgramsexplorer Click showMy DeviceCleaning TroubleshootingSymptoms Possible causes and Solutions Major Internal ComponentsCauses and solution keys Vacuum Degassing for the BDS Oxygen Sensor Oxygen Analyzer Maintenance Teledyne Analytical Instruments Specifications AppendixOperating Temperature 5-40 C Oxygen Analyzer AppendixRecommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountApplication notes Section I Product Identification Material Safety Data SheetSection II Physical and Chemical Data Section III -Physical Hazards Section IV Health Hazard DataSection VI Handling Information Alarms IndexOxygen Analyzer Index MsdsZero