Teledyne BDS-3960 operating instructions Installation

Page 38

Installation

BDS 3960

 

 

 

 

 

 

When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It will ignore your request to calibrate, and it will not remember that request.

Once a zero or span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the zero or span is complete, the calibration will repeat and the Cal Relay Contact (CRC) will close again.

For example:

1)Test the CRC. When the CRC is open, Send a zero command until the CRC closes (The CRC will quickly close.)

2)When the CRC closes, remove the zero command.

3)When CRC opens again, send a span command until the CRC closes. (The CRC will quickly close.)

4)When the CRC closes, remove the span command.

When CRC opens again, zero and span are done, and the sample is being analyzed.

Note: The Remote Valve connections (described below) provides signals to ensure that the zero and span gas valves will be controlled synchronously.

Range ID Relays: There are four dedicated Range ID relay contacts. The first three ranges are assigned to relays in ascending order—Low range is assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and High range is assigned to Range 3 ID. The fourth range is reserved for the over Range (100 ppm). Table 3-4 lists the pin connections. There is contact opening to indicate what range the analyzer is on. The contacts open when the analyzer is on that range.

Teledyne Analytical Instruments

38

Image 38
Contents Oxygen Analyzer Teledyne Analytical InstrumentsBDS Copyright 2002 Teledyne Analytical InstrumentsSpecific Model Information IiiSafety Messages Oxygen Analyzer Table of Contents Operation ViiIndex Teledyne Analytical Instruments MaintenanceAppendix ViiiList of Figures BDS-3960 Front PanelVacuum Degassing for the BDS Oxygen Sensor List of Tables Combustible GAS Usage XiiOverview Main Features of the AnalyzerIntroduction Typical ApplicationsFront Panel Operator Interface BDS-3960 Front Panel Remote Probe Rear Panel Equipment InterfaceAnalog Outputs Range ID ContactsModel BDS 3960 Rear Panel BDS Sensor Operational TheoryIntroduction Principles of OperationOxygen Analyzer Operational Theory Cross Section of the BDS Oxygen Sensor Gas Flow Rate BDS sensor output at different gas flow rateGas Pressure Temperature effectRecovery from High Level Oxygen Exposure Background Gas Compatibility Typical Purge-down Curve After Air SaturationMaintenance StabilitySample System Electronics and Signal Processing Flow DiagramOxygen Analyzer Operational Theory BDS 3960 Electronics Block Diagram Page Mounting the Analyzer InstallationUnpacking the Analyzer Oxygen Analyzer Installation Model BDS 3960 Front PanelRear Panel Connections Gas ConnectionsElectrical Connections Equipment Interface Connector Pin Arrangement Threshold Alarm Pin FunctionSystem Alarm Pin ContactCal Contact Installation Electrolyte Refill of BDS Sensor Testing the System Adding Electrolyte to the BDS SensorPowering Up the System Operation Auto Ranging onOxygen Analyzer Maintenance Analyzer applicationSystem Screen Communication Information and Calibration ParametersSetting Software Parameters Filter, Gas Factor, Tmp. Coeff Gas Factor Parameter Selection BoxTemperature Coefficient Parameter Default Values Calibration of the Analyzer Zero CalSpan Cal Alarms Function Maintenance Setting the Analog Output Ranges Range FunctionRange Options List Box Signal Output Fixed Range AnalysisSwitching The Program Back To Front Oxygen Analyzer Maintenance Maintenance Routine Maintenance Adding Water to the BDS SensorFuse Replacement Adding Water into the BDS sensorBattery Backup Replacement Removing Fuse Block from HousingReinstalling Application software to PPC Click startProgramsexplorer Click showMy DeviceSymptoms Possible causes and Solutions CleaningTroubleshooting Major Internal ComponentsCauses and solution keys Vacuum Degassing for the BDS Oxygen Sensor Oxygen Analyzer Maintenance Teledyne Analytical Instruments Specifications AppendixOperating Temperature 5-40 C Oxygen Analyzer AppendixRecommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountApplication notes Section II Physical and Chemical Data Material Safety Data SheetSection I Product Identification Section III -Physical Hazards Section IV Health Hazard DataSection VI Handling Information Alarms IndexOxygen Analyzer Index MsdsZero