Oxygen Analyzer | Maintenance | |
|
|
|
Filter | 5 minutes |
Gas Factor | 1.00 |
Temperature Coef. | 0 ppb/degree centigrade |
Alarm 1 setpoint | 1000 ppb |
Alarm 2 setpoint | 100 ppb |
Pressing the Set Defaults button brings up a confirmation dialog box. Press Yes to reset to the defaults or No to keep the current values.
4.4Calibration of the Analyzer
The analyzer must be calibrated prior to use. For most applications where the desired range of measurement is 0 to 10 ppm, or less we recommend the analyzer be calibrated using a span gas with a concentration between 7.0 to 9.0 ppm oxygen in nitrogen. This will require that calibration be performed in the
Before the cell is ready for calibration, it must be purged with sample gas to a low oxygen
4.4.1Zero Cal
The BDS Sensor has a zero offset of less than 5 ppb oxygen. Normally, the offset slowly decreases during the first 7 to 10 days of operation, and is expected to reach a steady value after this time.
Generally, the value of the zero offset is part of the oxygen reading of the sample gas as shown by the analyzer readout. As an example, a reading of 5 ppb oxygen may include 0.4 ppm oxygen in the sample gas and a 5 ppb zero offset.
The determination of the zero offset requires the use of oxygen free gas to the analyzer. The BDS 3960 has an oxygen scrubber as an integral part of its sample system. The software and electronics automatically
Teledyne Analytical Instruments | 49 |