Oxygen Analyzer | Maintenance | |
|
|
|
as
Note: Refer to Section 4.3.2.2 to find maximum working range.
4.6.2Fixed Range Analysis
The autoranging mode of the instrument can be overridden, forcing the analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, click on the list box of the main screen, as shown on figure 4.3, then select on the range of interest.
4.8Signal Output
The standard Model BDS 3960 Oxygen Analyzer is equipped with two
See Rear Panel in Chapter 3, Installation, for illustration and pin configuration of the interface connector.
The signal output for concentration is linear over the currently selected analysis range. For example, if the analyzer is set on a range that was defined as
Table
| Voltage Signal | Current Signal |
ppb O2 | Output (VDC) | Output (mA DC) |
0 | 0.0 | 4.0 |
10 | 0.1 | 5.6 |
20 | 0.2 | 7.2 |
30 | 0.3 | 8.8 |
40 | 0.4 | 10.4 |
50 | 0.5 | 12.0 |
Teledyne Analytical Instruments | 55 |