Teledyne BDS-3960 Operational Theory, Introduction, BDS Sensor, Principles of Operation

Page 18

Operational Theory

BDS 3960

 

 

 

 

 

 

Operational Theory

2.1Introduction

The analyzer is composed of four subsystems:

BDS Sensor

Sample System

Pocket PC

Electronic I/O Signal Processing and Display

The sample system is designed to accept the sample gas and

transport it through the analyzer without contaminating or altering the sample prior to analysis. The BDS Sensor is an electrochemical device that translates the amount of oxygen present in the sample into an electrical current. The Pocket PC processes the sensor signal and sends messages to the I/O electronics to correctly display oxygen value as well as control other signals to the customer interface. The Electronic I/O signal processing amplifies the sensor signal, digitizes the sensor reading and sends them to the pocket PC for processing. Then it receives commands from the pocket PC to manipulate signals for the customer interface.

2.2BDS Sensor

2.2.1 Principles of Operation

The BDS oxygen sensor technology developed at Teledyne Analytical Instruments is a result of TAI’s heavy investment on R&D and expertise established during the half-century’s manufacturing of electrochemical oxygen sensor. It stands for Bipotentiostat Driven Sensor. A BDS oxygen sensor accurately translates the oxygen level in the sample gas into to an electrical current signal.

A potentiostat contains three electrodes: a working electrode, a reference electrode and a counter electrode. A Bipotentiostat is a combination of two potentiostats that share the reference electrode and the counter electrode. The potential at the working electrode is precisely

Teledyne Analytical Instruments

18

Image 18
Contents Oxygen Analyzer Teledyne Analytical InstrumentsBDS Copyright 2002 Teledyne Analytical InstrumentsSpecific Model Information IiiSafety Messages Oxygen Analyzer Table of Contents Operation ViiIndex Teledyne Analytical Instruments MaintenanceAppendix ViiiList of Figures BDS-3960 Front PanelVacuum Degassing for the BDS Oxygen Sensor List of Tables Combustible GAS Usage XiiOverview Main Features of the AnalyzerIntroduction Typical ApplicationsFront Panel Operator Interface BDS-3960 Front Panel Remote Probe Rear Panel Equipment InterfaceAnalog Outputs Range ID ContactsModel BDS 3960 Rear Panel BDS Sensor Operational TheoryIntroduction Principles of OperationOxygen Analyzer Operational Theory Cross Section of the BDS Oxygen Sensor Gas Flow Rate BDS sensor output at different gas flow rateTemperature effect Recovery from High Level Oxygen ExposureGas Pressure Background Gas Compatibility Typical Purge-down Curve After Air SaturationMaintenance StabilitySample System Electronics and Signal Processing Flow DiagramOxygen Analyzer Operational Theory BDS 3960 Electronics Block Diagram Page Installation Unpacking the AnalyzerMounting the Analyzer Oxygen Analyzer Installation Model BDS 3960 Front PanelRear Panel Connections Gas ConnectionsElectrical Connections Equipment Interface Connector Pin Arrangement Threshold Alarm Pin FunctionSystem Alarm Pin ContactCal Contact Installation Electrolyte Refill of BDS Sensor Testing the System Adding Electrolyte to the BDS SensorPowering Up the System Operation Auto Ranging onOxygen Analyzer Maintenance Analyzer applicationSystem Screen Communication Information and Calibration ParametersSetting Software Parameters Filter, Gas Factor, Tmp. Coeff Gas Factor Parameter Selection BoxTemperature Coefficient Parameter Default Values Calibration of the Analyzer Zero CalSpan Cal Alarms Function Maintenance Setting the Analog Output Ranges Range FunctionRange Options List Box Signal Output Fixed Range AnalysisSwitching The Program Back To Front Oxygen Analyzer Maintenance Maintenance Routine Maintenance Adding Water to the BDS SensorFuse Replacement Adding Water into the BDS sensorBattery Backup Replacement Removing Fuse Block from HousingReinstalling Application software to PPC Click startProgramsexplorer Click showMy DeviceSymptoms Possible causes and Solutions CleaningTroubleshooting Major Internal ComponentsCauses and solution keys Vacuum Degassing for the BDS Oxygen Sensor Oxygen Analyzer Maintenance Teledyne Analytical Instruments Specifications AppendixOperating Temperature 5-40 C Oxygen Analyzer AppendixRecommended 2-Year Spare Parts List Qty Part Number DescriptionDrawing List 19-inch Relay Rack Panel MountApplication notes Material Safety Data Sheet Section I Product IdentificationSection II Physical and Chemical Data Section III -Physical Hazards Section IV Health Hazard DataSection VI Handling Information Alarms IndexOxygen Analyzer Index MsdsZero