Teledyne BDS-3960 operating instructions Temperature Coefficient

Page 47

Oxygen Analyzer

Maintenance

 

 

 

*Output normalized by calibrating with a span gas in N2.

The values in Table 4-1 are theoretical but should give a first approximation as to what the output of the sensor will be in the new background gas. The BDS 3960 will make further refinements once the proper gas factor is input.

The working range of adjustment is 0.25 to 2.50. This factor will divide the output. For example if the factor is set to 2.00, the output of the sensor, when read by the PPC application will be divided by two.

Special consideration on the working range: Changing the gas correction factor has an effect on the maximum working range of the analyzer, e.g.: if a gas factor of 2.00 is selected the maximum working range of the analyzer is 50 ppm. Any reading above this, may saturate the amplifier.

4.3.2.3TEMPERATURE COEFFICIENT

The output of the sensor is affected by temperature. There are two types of temperature compensation on the BDS 3960. One is the compensation to high levels of oxygen, more than 100 ppb. This effect is transparent to the user and is handled by the software application on the PPC. It requires no input from the user. The second compensation involves the baseline drift due to temperature. This value is generally set at the factory but can be modified or adjusted by the user. Since the sensor is temperature controlled, the default is zero. The following discussion is just information on its possibilities.

The baseline drift temperature coefficient is a number with units of ppb/degree centigrade and must be matched to the sensor’s characteristic drift over temperature. This coefficient ranges from 0.50 to 1.75 ppb/°C.

Note: The coefficient is different from sensor to sensor. If the sensor is replaced, a new coefficient must be entered. TAI can supply this coefficient or it may be determined by the user.

The coefficient can be adjusted between 0.00 and 5.00 using the Up and Down buttons on the instrument. To estimate it in the field:

1.Set the coefficient to zero.

2.Run the analyzer on “Zero” calibration gas for two weeks or until a baseline stability is reached, i.e. the oxygen reading does not fluctuate.

Teledyne Analytical Instruments

47

Image 47
Contents Teledyne Analytical Instruments Oxygen AnalyzerCopyright 2002 Teledyne Analytical Instruments BDSIii Specific Model InformationSafety Messages Oxygen Analyzer Table of Contents Vii OperationViii MaintenanceAppendix Index Teledyne Analytical InstrumentsBDS-3960 Front Panel List of FiguresVacuum Degassing for the BDS Oxygen Sensor List of Tables Xii Combustible GAS UsageTypical Applications Main Features of the AnalyzerIntroduction OverviewFront Panel Operator Interface BDS-3960 Front Panel Range ID Contacts Rear Panel Equipment InterfaceAnalog Outputs Remote ProbeModel BDS 3960 Rear Panel Principles of Operation Operational TheoryIntroduction BDS SensorOxygen Analyzer Operational Theory Cross Section of the BDS Oxygen Sensor BDS sensor output at different gas flow rate Gas Flow RateGas Pressure Temperature effectRecovery from High Level Oxygen Exposure Typical Purge-down Curve After Air Saturation Background Gas CompatibilityStability MaintenanceSample System Flow Diagram Electronics and Signal ProcessingOxygen Analyzer Operational Theory BDS 3960 Electronics Block Diagram Page Mounting the Analyzer InstallationUnpacking the Analyzer Model BDS 3960 Front Panel Oxygen Analyzer InstallationGas Connections Rear Panel ConnectionsElectrical Connections Equipment Interface Connector Pin Arrangement Pin Function Threshold AlarmPin Contact System AlarmCal Contact Installation Electrolyte Refill of BDS Sensor Adding Electrolyte to the BDS Sensor Testing the SystemPowering Up the System Auto Ranging on OperationAnalyzer application Oxygen Analyzer MaintenanceCommunication Information and Calibration Parameters System ScreenSetting Software Parameters Filter, Gas Factor, Tmp. Coeff Parameter Selection Box Gas FactorTemperature Coefficient Parameter Default Values Zero Cal Calibration of the AnalyzerSpan Cal Alarms Function Maintenance Range Function Setting the Analog Output RangesRange Options List Box Fixed Range Analysis Signal OutputSwitching The Program Back To Front Oxygen Analyzer Maintenance Adding Water to the BDS Sensor Maintenance Routine MaintenanceAdding Water into the BDS sensor Fuse ReplacementRemoving Fuse Block from Housing Battery Backup ReplacementClick startProgramsexplorer Click showMy Device Reinstalling Application software to PPCMajor Internal Components CleaningTroubleshooting Symptoms Possible causes and SolutionsCauses and solution keys Vacuum Degassing for the BDS Oxygen Sensor Oxygen Analyzer Maintenance Teledyne Analytical Instruments Appendix SpecificationsOxygen Analyzer Appendix Operating Temperature 5-40 CQty Part Number Description Recommended 2-Year Spare Parts List19-inch Relay Rack Panel Mount Drawing ListApplication notes Section II Physical and Chemical Data Material Safety Data SheetSection I Product Identification Section IV Health Hazard Data Section III -Physical HazardsSection VI Handling Information Index AlarmsMsds Oxygen Analyzer IndexZero